![GE Baker Hughes oxy.IQ User Manual Download Page 17](http://html.mh-extra.com/html/ge/baker-hughes-oxy-iq/baker-hughes-oxy-iq_user-manual_18634017.webp)
Chapter 2. Installation
oxy.IQ User’s Manual
7
2.1 Mounting the oxy.IQ (cont.)
IMPORTANT:
The maximum operating pressure for the oxy.IQ is 10
psi, and the burst pressure of the unit is 200 psi. Be sure
the sample conditioning system is designed to maintain
the oxy.IQ pressure below these limits, and that the
oxy.IQ outlet is vented to atmosphere during operation
and calibration.
3.
Using PTFE tape as a sealant, connect the sample gas inlet and
outlet to the 1/8” NPT ports on the sensor manifold (see
below). Either port may be used as the inlet or the outlet, as the
direction of flow does not matter.
Figure 4: Sensor Manifold Installation
Sample Inlet
Sample Outlet
Sensor Manifold
PTFE Tape
Summary of Contents for Baker Hughes oxy.IQ
Page 1: ...Oxygen 910 296 Rev E July 2018 bhge com oxy IQ Panametrics Oxygen Analyzer User s Manual ...
Page 2: ......
Page 4: ...ii oxy IQ User s Manual no content intended for this page ...
Page 30: ...Chapter 3 Initial Setup Operation 20 oxy IQ User s Manual no content intended for this page ...
Page 38: ...Chapter 4 User Programming 28 oxy IQ User s Manual no content intended for this page ...
Page 46: ...Chapter 6 Specifications 36 oxy IQ User s Manual no content intended for this page ...
Page 58: ...Appendix B Menu Maps 48 oxy IQ User s Manual no content intended for this page ...
Page 60: ...Appendix C Order String 50 oxy IQ User s Manual no content intended for this page ...
Page 66: ...Appendix D Certifications 56 oxy IQ User s Manual D 2 ATEX IECEx MAM Ex Certificate ...
Page 67: ...Appendix D Certifications oxy IQ User s Manual 57 D 2 ATEX IECEx MAM Ex Certificate cont ...
Page 74: ...Appendix D Certifications 64 oxy IQ User s Manual D 5 IECEx Certificate of Conformity ...
Page 75: ...Appendix D Certifications oxy IQ User s Manual 65 D 5 IECEx Certificate of Conformity cont ...
Page 76: ...Appendix D Certifications 66 oxy IQ User s Manual D 5 IECEx Certificate of Conformity cont ...
Page 77: ...Appendix D Certifications oxy IQ User s Manual 67 D 5 IECEx Certificate of Conformity cont ...
Page 87: ...Addendum A oxy IQ Safety Manual 77 oxy IQ User s Manual no content intended for this page ...
Page 89: ... no content intended for this page ...
Page 90: ......