The absolute accuracy of the instrument will be depen-
dent on the accuracy of the user reference blocks and
will no longer be to the accuracy stated in the “Speci-
fi cations” chapter of these instructions. It is not rec-
ommended that measurements are made outside the
range of the two sample values; as the accuracy will
be totally dependent on the absolute accuracy of the
user reference blocks.
For the relative reading accuracy to be the same as
for GE Inspection Technologies STD blocks, the two
user reference block values must be accurate and lie
either side of the measurement range. e.g. If the mea-
surement range is to be between 40 and 45 % IACS
then STD 1 should ideally be 50 % IACS or more and
STD 2 should be 35 % IACS or less.
It is possible to set the AutoSigma 3000 DL by using
one user reference block and one STD block, provided
that the above rules are complied with, the absolute
accuracy will again be dependent on the accuracy of
the user reference block.
5.5
Probe setting (SET PROBE)
Before the probe setting on the reference blocks can
take place, the value of the blocks must be entered in
the STD 1 and STD 2 menu slots in the units selected
(i.e. IACS or MS/m) The probe code must also be ap-
propriate to the probe in use. Proceed following sec-
tion 5.4 to set the reference block values. Then with
the instrument on and in the Measurement mode:
Entering reference block values
Operation
5 - 6
Issue 04, 06/2007
AutoSigma 3000
Summary of Contents for AutoSigma 3000 DL
Page 1: ...AutoSigma 3000 Technical Reference and Operation Manual Ident No 071 002 022 47DH01 105 952...
Page 6: ...0 6 Issue 04 06 2007 AutoSigma 3000...
Page 7: ...Introduction 1 AutoSigma 3000 Issue 04 06 2007 1 1...
Page 12: ...1 6 Issue 04 06 2007 AutoSigma 3000...
Page 13: ...Standard package and accessories 2 AutoSigma 3000 Issue 04 06 2007 2 1...
Page 16: ...2 4 Issue 04 06 2007 AutoSigma 3000...
Page 17: ...Preparation for operation 3 AutoSigma 3000 Issue 04 06 2007 3 1...
Page 27: ...Basic functions 4 AutoSigma 3000 Issue 04 06 2007 4 1...
Page 31: ...Operation 5 AutoSigma 3000 Issue 04 06 2007 5 1...
Page 47: ...External data handling devices 6 AutoSigma 3000 Issue 04 06 2007 6 1...
Page 52: ...6 6 Issue 04 06 2007 AutoSigma 3000...
Page 53: ...Specifications 7 AutoSigma 3000 Issue 04 06 2007 7 1...
Page 57: ...Maintenance and care 8 AutoSigma 3000 Issue 04 06 2007 8 1...
Page 59: ...Appendix 9 AutoSigma 3000 Issue 04 06 2007 9 1...
Page 62: ...9 4 Issue 04 06 2007 AutoSigma 3000...
Page 63: ...Index 10 AutoSigma 3000 Issue 04 06 2007 10 1...