Using the brake
The central brake holds the system in place.
Do not use the brake while moving the anesthesia
system. This could cause the machine to tip over. Only
use the brake to keep the system in place.
AC.22.009
1.
Push down on the brake pedal to lock the system in place.
2.
Lift up on the brake pedal to release the brake.
Using the O2 flush button
The O2 flush button supplies a high flow of O2 to the breathing
system.
1.
Push the O2 flush button to deliver a high flow of O2.
2.
Release the O2 flush button to stop the delivery of a high flow of
O2.
Positioning the display
The display can be moved for optimal viewing.
1.
Unlock the display arm.
2.
Raise or lower the display arm to adjust the height of the
display.
3.
Lock the display arm.
4.
Rotate the display arm toward or away from the system to adjust
the horizontal position of the display.
5.
Tilt the display up or down to adjust the vertical angle of the
display.
WARNING
Aisys CS²
2-4
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...