Inspection

In order to assure continued performance of this x

ray equipment, a periodic inspection program must be

established.  It is the owner’s responsibility to supply or arrange for this service.
Periodic Maintenance can locate potential problems before they occur. Since maintenance checks involve cost, it
is not practical to attempt to assure zero failures. The following inspection requirements, with suggested
inspection intervals, are GE Healthcare recommendation for the most cost effective maintenance schedule.
Inspection intervals are based on average daily use of one eight hour shift. More frequent inspection is appropriate
where equipment use is above average. Equipment that regularly exhibits out of tolerance conditions should also
be inspected more frequently.

1-5

HHS Testing

The United States Department of Health and Human Services (HHS) has established performance requirements
for diagnostic x

ray equipment. These requirements are defined in Title 21 of the Code of federal regulations (21

CFR), and apply only to certain specified components identified as “certified equipment.”
The manufacturer of specified diagnostic x

ray components must certify that the components:

1. Perform as required by the HHS standard when installed, adjusted, and tested as specified in the

manufacturer ’s instructions to the assembler.
Refer to Direction 2173222

100 

AMX

4+ Installation

 for HHS compliance testing as it pertains to the

AMX

4+ during installation.

2. Will continue to comply when maintained in accordance with the manufacturer’s instructions.

The importance of HHS compliance testing cannot be over emphasized.  Before starting Periodic
Maintenance procedures, review Direction 46

013894 

System Field Test for HHS. 

Pay particular attention to

“Introduction” (Tab 1 of Direction 46

013894).

1-6

Tools and Materials

In addition to the standard service representatives tool kit, the following items are required:

Direction 2173221

100, 

AMX

4+ Ratings & Specifications 

and associated tools and materials.

Direction 2173224

100, 

AMX

4+ Functional Check 

and associated tools and materials.

Direction 2173225

100, 

AMX

4+ Service  

and associated tools and materials.

Direction 2190775

100, 

Stranded Steel Cable Inspection

, and associated tools and materials

SECTION 2 

 PM SCHEDULES

Your AMX 4+ does not require any planned maintenance to be performed for a 13 month period following
installation. PM should begin in the 13th month that follows your system’s installation anniversary date. 
Following the 13th month, PMs should be performed at the frequency indicated in Section 3 that follows. 

Note:

The hand switch 

MUST 

be replaced at least once in every 3 years.

Summary of Contents for AMX 4+

Page 1: ...GE Healthcare AMX 4 Periodic Maintenance Models 2169360 2236420 and 2275938 Direction 2173227 100 Revision 10 Copyright 1996 c by General Electric Company All rights Reserved...

Page 2: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 ii THIS PAGE INTENTIONALLY LEFT BLANK...

Page 3: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 iii...

Page 4: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 iv...

Page 5: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 v...

Page 6: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 vi...

Page 7: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 vii...

Page 8: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 viii...

Page 9: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 ix...

Page 10: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 x...

Page 11: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 xi...

Page 12: ...s sold with the understanding that the General Electric Company Medical Systems Group its agents and representatives have no responsibility for injury or damage which may result from improper use of t...

Page 13: ...Electric representative or a hospital receiving agent Whether noted or concealed damage MUST be reported to the carrier immediately upon discovery or in any event within 14 days after receipt and the...

Page 14: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 xiv THIS PAGE INTENTIONALLY LEFT BLANK...

Page 15: ...dded note about installing local language label over English rating plate 7 Feb 15 2008 Updated Table 1 2 to remove old labels and add new international labels Updated flowchart in Table 2 1 to modify...

Page 16: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 xvi THIS PAGE INTENTIONALLY LEFT BLANK...

Page 17: ...0 2236420 2275938 REV 10 DIRECTION 2173227 100 xvii TABLE OF CONTENTS SECTION 1 INTRODUCTION 1 1 1 Identification 1 1 2 General 1 1 3 Cleaning 2 1 4 Inspection 2 1 5 HHS Testing 2 1 6 Tools and Materi...

Page 18: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 xviii THIS PAGE INTENTIONALLY LEFT BLANK...

Page 19: ...JH domestic aec tech switch 2169360 10 A0659FB 2236420 10 2275938 10 A0659JJ iec emc 2169360 A0659A 2236420 2275938 A0659J iec emc aec 2169360 2 A0659AA 2236420 2 2275938 2 A0659JA iec emc tech switch...

Page 20: ...althcare representative will be glad to discuss this plan 1 3 Cleaning The AMX 4 should be cleaned frequently particularly if corroding chemicals are present Use a cloth slightly moistened in warm soa...

Page 21: ...fy that the components 1 Perform as required by the HHS standard when installed adjusted and tested as specified in the manufacturer s instructions to the assembler Refer to Direction 2173222 100 AMX...

Page 22: ...all applicable checks to the main safety and follower counterweight cables Exception is that this is a 12 month interval for the AMX 4 HHS TESTS Excluding AEC Refer to HHS Tests Excluding AEC in Sect...

Page 23: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 5 ILLUSTRATION 2 2 DRIVE BELT...

Page 24: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 6 THIS PAGE INTENTIONALLY LEFT BLANK...

Page 25: ......

Page 26: ...General Electric Company GE Medical Systems a General Electric Company going to market as GE Healthcare 3000 N Grandview Boulevard Waukesha Wisconsin 53188 USA www gehealthcare com...

Reviews: