4+ PERIODIC MAINTENANCE

GE Healthcare

MODELS: 2169360, 2236420 & 2275938

REV 

10

DIRECTION 2173227

100

4

SECTION 3 

 INSPECTION

The recommended interval between maintenance checks is 12 months.  The following flowchart shows the
recommended sequence for performing an annual periodic maintenance check on the AMX

4+.

ILLUSTRATION 2

1

INSPECTION STEERING

AEC HHS TESTS

Refer to ”AEC HHS Tests” in Section 3 of

Direction 2173221

100, 

AMX

4+ Ratings &

Specifications

, and perform all defined

procedures.

Mobil

AID AEC

Present?

YES

NO

Done

FUNCTIONAL CHECK

Refer to ”Functional Check” in Direction

2173224

100, 

AMX

4+ Functional Check

,

and perform all defined checks.

POWER CORD CHECK

Refer to ”Power Cord Check” in Direction

2173224

100, 

AMX

4+ Functional Check

,

and perform all defined checks.

COUNTERWEIGHT CABLE

Refer to Direction 2190775

100, 

Stranded

Steel Cable Inspection

, and perform all

applicable checks to the main, safety, and

follower counterweight cables.  Exception is

that this is a 12 month interval for the

AMX

4+.

HHS TESTS (Excluding AEC)

 Refer to ”HHS Tests (Excluding AEC)” in

Section 3 of Direction 2173221

100,

AMX

4+ Ratings & Specifications

, and

perform all defined procedures.

COLLIMATOR INTERFACE TIGHTNESS

Verify that there is no movement between 

the collimator and X−ray tube.  If there is 

no movement, verify mounting screws 

tightness for torque specs per Direction 

2173225-100. If any movement is 

detected, you 

MUST

 remove the 

collimator and inspect all related fasteners.  

Refer to ”Servicing the Collimator” in 

Direction 2173225−100, 

AMX−4+ Service

for more information.

DRIVE BELT TENSION

Refer to Illustration 2

2. Adjust the tension

 

of drive belts on both wheels:

Loosen the bolt and nut P62 (about one

 

turn) to allow the spring P56 to apply the

 

tension to the drive belts. Then tighten the

 

nut and torque to 50

60 ft.

lbs to maintain

 

the position of the motor assembly. Finally,

 

apply Loctite 242 P99 on the nut to keep the

 

fastening from loosening.

Summary of Contents for AMX 4+

Page 1: ...GE Healthcare AMX 4 Periodic Maintenance Models 2169360 2236420 and 2275938 Direction 2173227 100 Revision 10 Copyright 1996 c by General Electric Company All rights Reserved...

Page 2: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 ii THIS PAGE INTENTIONALLY LEFT BLANK...

Page 3: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 iii...

Page 4: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 iv...

Page 5: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 v...

Page 6: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 vi...

Page 7: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 vii...

Page 8: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 viii...

Page 9: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 ix...

Page 10: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 x...

Page 11: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 xi...

Page 12: ...s sold with the understanding that the General Electric Company Medical Systems Group its agents and representatives have no responsibility for injury or damage which may result from improper use of t...

Page 13: ...Electric representative or a hospital receiving agent Whether noted or concealed damage MUST be reported to the carrier immediately upon discovery or in any event within 14 days after receipt and the...

Page 14: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 xiv THIS PAGE INTENTIONALLY LEFT BLANK...

Page 15: ...dded note about installing local language label over English rating plate 7 Feb 15 2008 Updated Table 1 2 to remove old labels and add new international labels Updated flowchart in Table 2 1 to modify...

Page 16: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 xvi THIS PAGE INTENTIONALLY LEFT BLANK...

Page 17: ...0 2236420 2275938 REV 10 DIRECTION 2173227 100 xvii TABLE OF CONTENTS SECTION 1 INTRODUCTION 1 1 1 Identification 1 1 2 General 1 1 3 Cleaning 2 1 4 Inspection 2 1 5 HHS Testing 2 1 6 Tools and Materi...

Page 18: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 xviii THIS PAGE INTENTIONALLY LEFT BLANK...

Page 19: ...JH domestic aec tech switch 2169360 10 A0659FB 2236420 10 2275938 10 A0659JJ iec emc 2169360 A0659A 2236420 2275938 A0659J iec emc aec 2169360 2 A0659AA 2236420 2 2275938 2 A0659JA iec emc tech switch...

Page 20: ...althcare representative will be glad to discuss this plan 1 3 Cleaning The AMX 4 should be cleaned frequently particularly if corroding chemicals are present Use a cloth slightly moistened in warm soa...

Page 21: ...fy that the components 1 Perform as required by the HHS standard when installed adjusted and tested as specified in the manufacturer s instructions to the assembler Refer to Direction 2173222 100 AMX...

Page 22: ...all applicable checks to the main safety and follower counterweight cables Exception is that this is a 12 month interval for the AMX 4 HHS TESTS Excluding AEC Refer to HHS Tests Excluding AEC in Sect...

Page 23: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 5 ILLUSTRATION 2 2 DRIVE BELT...

Page 24: ...AMX 4 PERIODIC MAINTENANCE GE Healthcare MODELS 2169360 2236420 2275938 REV 10 DIRECTION 2173227 100 6 THIS PAGE INTENTIONALLY LEFT BLANK...

Page 25: ......

Page 26: ...General Electric Company GE Medical Systems a General Electric Company going to market as GE Healthcare 3000 N Grandview Boulevard Waukesha Wisconsin 53188 USA www gehealthcare com...

Reviews: