4.1.2
Room requirements
Introduction
This section describes the requirements for the transportation route and the room where
the ÄKTA pure instrument is placed.
WARNING
Protective ground. The ÄKTA pure instrument must always be
connected to a grounded power outlet.
WARNING
Only use grounded power cords delivered or approved by GE.
WARNING
Do not block access to the power switch and power cord. The
power switch must always be easy to access. The power cord with
plug must always be easy to disconnect.
CAUTION
Explosion hazard. To avoid building up an explosive atmosphere
when using flammable liquids, make sure that the room ventilation
meets the local requirements.
Transportation route
Doors, corridors and elevators must have a minimum width of 65 cm to allow for trans-
porting the instrument. Allow additional space for moving around corners.
ÄKTA pure Operating Instructions 29022997 AF
59
4 Installation
4.1 Site preparation
4.1.2 Room requirements