GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 3
Direction 18000
8-6-5
2) BREAKOUT POLLING
Description:
(SB software does not use Breakout Polling in the UPAP process. Rev. “SC” does).
Breakout polling is a software method to resynchronize the RS232 communications port (6809) to the HIB.
An out-of-sync condition is rare but can occur if the AP was shut down abnormally. An example is a
“CNTL A” during the UPAP serial transfer section of deadstarts. When out-of-sync, the 6809 communications
processor is still looking for more data on the link - it takes any serial transfer and places it into public
memory as if it were a continuation of the data string - it cannot distinguish between commands or data. This
mode will continue until the internal byte counter is satisfied and the data transfer is complete.
Error conditions:
More than 1 BREAKOUT POLLING sequence signifies a problem. RAP’s in WTERM will also fail, and failure
points are the same. Here is a sample error message:
BREAKOUT POLLING 6809
ASSUMING 6809 IN BYTE COUNT LOOP
BREAKOUT POLLING 6809
ASSUMING 6809 IN BYTE COUNT LOOP
BREAKOUT POLLING 6809
ASSUMING 6809 IN BYTE COUNT LOOP
The failure points are:
a) Cable break
b) Loss of
±
12 volts at drivers in AP - IOC or DG - HIB
c ) Loss of + 5v at AP
d) Wrong baud rate settings on IOC card (DIP switch)
e) Wrong baud rate setting or device select on HIB (DIP switches)
f)
Bad RS232 driver / receiver IC’s on IOC or HIB
g) Failure of HIB or IOC
3) HARDWARE RESET (1st)
Description:
The first section of deadstart reset contains two parts. The first is a simple RESET that contains the same
primitive serial transfer protocol as a RAP command from WTERM. The second part is a serial transfer that
contains data. This transfer requires the use of the MASTER card, public memory, and the public memory
bus. The contents of this secondary transfer is what kicks off the MASTER and SLAVE diagnostics (that
reside in the AP PROM) to continue the deadstart.
Error conditions:
a) Error message seen from simple RESET failure:
*E* Transfer did not succeed
Summary of Contents for Advantium CT9800
Page 2: ......
Page 49: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 5 ILLUSTRATION 4 3 3 ...
Page 122: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 4 9 6 ...
Page 138: ...Page Left Intentionally Blank ...
Page 227: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 3 ILLUSTRATION 6 9 2 ...
Page 234: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 10 ILLUSTRATION 6 9 8 ...
Page 240: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 16 ILLUSTRATION 6 9 12 ...
Page 315: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 1 1 ...
Page 316: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 1 2 ...
Page 317: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 1 3 ...
Page 319: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 1 ...
Page 320: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 2 ...
Page 321: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 3 ...
Page 322: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 4 ...
Page 323: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 5 ...
Page 324: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 6 ...
Page 325: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 7 ...
Page 326: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 8 ...
Page 327: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 9 ...
Page 328: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 10 ...
Page 329: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 11 ...
Page 330: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 12 ...
Page 331: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 13 ...
Page 332: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 7 2 14 ...
Page 333: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 14A ...
Page 335: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 7 2 15 ...
Page 336: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 7 2 16 ...
Page 337: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 7 2 17 ...
Page 338: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 18 ...
Page 339: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 19 ...
Page 340: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 20 ...
Page 341: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 7 2 21 ...
Page 342: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 22 ...
Page 343: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 23 ...
Page 344: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 24 ...
Page 345: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 25 ...
Page 346: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 26 ...
Page 347: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 27 ...
Page 348: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 28 37 46 218868G1 ...
Page 349: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 29 ...
Page 350: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 30 ...
Page 351: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 31 ...
Page 352: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 32 ...
Page 353: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 33 ...
Page 354: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 34 ...
Page 355: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 10 Direction 18000 7 2 35 ...
Page 356: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 36 ...
Page 357: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37 ...
Page 358: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37A ...
Page 360: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 38 ...
Page 361: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 39 ...
Page 362: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 40 ...
Page 363: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 41 ...
Page 364: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 7 2 42 ...
Page 365: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 43 ...
Page 366: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 44 ...
Page 367: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 45 ...
Page 368: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 46 ...
Page 369: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 47 ...
Page 370: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 47A ...
Page 372: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48 ...
Page 373: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48A ...
Page 375: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 49 ...
Page 376: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 50 ...
Page 377: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 51 ...
Page 379: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 3 1 ...
Page 381: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 4 1 ...
Page 382: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 4 2 ...
Page 383: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 4 3 ...
Page 384: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 7 4 4 ...
Page 411: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 3 5 ...
Page 487: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 3 ...
Page 489: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 5 ...
Page 490: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 6 ...
Page 491: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 7 ...
Page 492: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 8 ...
Page 493: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 9 ...
Page 494: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 10 ...
Page 495: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 11 ...
Page 496: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 12 ...
Page 497: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 13 ...
Page 498: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 14 ...
Page 499: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 15 ...
Page 500: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 16 ...
Page 501: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 17 ...
Page 502: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 18 ...
Page 503: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 19 ...
Page 504: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 20 ...
Page 506: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 22 ...
Page 507: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 23 ...
Page 508: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 24 ...
Page 509: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 25 ...
Page 510: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 26 ...