GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 12
Direction 18000
6-9A-24
4.
Review CBF projection
•
CBF must equal Isocenter
±
0.20 channels.
•
If OK, this procedure is done, otherwise proceed to Step 5.
5.
At gantry
•
Check for 0
°
tilt and table at load home normal
•
Remove side covers
•
Turn off scan drive switch
•
Flip over top cover
•
Swing out front cover
•
Rotate gantry to 270
°
(Collimator at 3:00 O’clock).
•
Install dial indicator and shift collimator the specified amount per Step 3. Refer to Section 9A-2-2 for
collimator positioning.
•
Close front cover
•
Arm scan drive switch
6.
Call Data Collection (DC) program
•
(DC)
Automatic down load
•
Repeat CBF scan, Step 2
9A-9 System Sag
The sag program measures the change in alignment during a rotating scan. It reports the peak amplitude
difference in channel numbers as well as the hysteresis or difference in channel numbers between first and last
views of a scan.
Refer to Illustration 6-9A-15 for the procedure flow.
1.
Use the calibration scan ISOCAL from Section 9A-7, Step 1.
2.
Use the CBF scan from Section 9A-8, Step 2.
3
Analyze the SAG run using the alignment program.
Enter CTDS
•
(AL)
Alignment
•
(3 )
SAG
•
(xx)
Select Exam # recorded in step 2 for CBF.
•
(yy)
Enter ISOCAL for calibration filename.
The program will calculate and display the maximum channel difference, the azimuthal location of the
channel difference, and the difference between first and last views in channels. Record the information
and then exit the Alignment program.
Summary of Contents for Advantium CT9800
Page 2: ......
Page 49: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 5 ILLUSTRATION 4 3 3 ...
Page 122: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 4 9 6 ...
Page 138: ...Page Left Intentionally Blank ...
Page 227: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 3 ILLUSTRATION 6 9 2 ...
Page 234: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 10 ILLUSTRATION 6 9 8 ...
Page 240: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 16 ILLUSTRATION 6 9 12 ...
Page 315: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 1 1 ...
Page 316: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 1 2 ...
Page 317: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 1 3 ...
Page 319: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 1 ...
Page 320: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 2 ...
Page 321: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 3 ...
Page 322: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 4 ...
Page 323: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 5 ...
Page 324: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 6 ...
Page 325: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 7 ...
Page 326: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 8 ...
Page 327: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 9 ...
Page 328: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 10 ...
Page 329: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 11 ...
Page 330: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 12 ...
Page 331: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 13 ...
Page 332: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 7 2 14 ...
Page 333: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 14A ...
Page 335: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 7 2 15 ...
Page 336: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 7 2 16 ...
Page 337: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 7 2 17 ...
Page 338: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 18 ...
Page 339: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 19 ...
Page 340: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 20 ...
Page 341: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 7 2 21 ...
Page 342: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 22 ...
Page 343: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 23 ...
Page 344: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 24 ...
Page 345: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 25 ...
Page 346: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 26 ...
Page 347: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 27 ...
Page 348: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 28 37 46 218868G1 ...
Page 349: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 29 ...
Page 350: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 30 ...
Page 351: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 31 ...
Page 352: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 32 ...
Page 353: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 33 ...
Page 354: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 34 ...
Page 355: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 10 Direction 18000 7 2 35 ...
Page 356: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 36 ...
Page 357: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37 ...
Page 358: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37A ...
Page 360: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 38 ...
Page 361: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 39 ...
Page 362: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 40 ...
Page 363: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 41 ...
Page 364: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 7 2 42 ...
Page 365: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 43 ...
Page 366: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 44 ...
Page 367: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 45 ...
Page 368: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 46 ...
Page 369: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 47 ...
Page 370: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 47A ...
Page 372: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48 ...
Page 373: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48A ...
Page 375: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 49 ...
Page 376: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 50 ...
Page 377: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 51 ...
Page 379: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 3 1 ...
Page 381: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 4 1 ...
Page 382: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 4 2 ...
Page 383: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 4 3 ...
Page 384: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 7 4 4 ...
Page 411: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 3 5 ...
Page 487: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 3 ...
Page 489: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 5 ...
Page 490: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 6 ...
Page 491: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 7 ...
Page 492: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 8 ...
Page 493: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 9 ...
Page 494: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 10 ...
Page 495: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 11 ...
Page 496: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 12 ...
Page 497: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 13 ...
Page 498: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 14 ...
Page 499: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 15 ...
Page 500: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 16 ...
Page 501: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 17 ...
Page 502: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 18 ...
Page 503: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 19 ...
Page 504: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 20 ...
Page 506: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 22 ...
Page 507: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 23 ...
Page 508: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 24 ...
Page 509: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 25 ...
Page 510: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 26 ...