9100c
4-4
M1207026
4.2 Service Mode
The
Service Mode
is used to test, calibrate, or troubleshoot ventilator
related components in the anesthesia machine.
The way to enter the service mode is:
•
With the system switch in Standby, push and hold in the
adjustment knob while setting the system switch to On. Hold
the adjustment knob pushed in until the “Operation Select”
menu appears. Use the adjustment knob to highlight “Service
Mode”, then push the adjustment knob to confirm the
selection.
From this menu you can:
•
Go to the
Service Mode
•
Return to
Normal Mode
Service Mode Menu
The
Service Mode
main menu displays the non-selectable ventilator
information and the service tests you can select.
The selectable service tests are displayed in categorical order. But
you can select the service test from this menu in any order. The
following sections in this manual are sequenced in the order that they
appear on the screen.