
9 Repair Procedure
M1207026
9-3
WARNING
To prevent fires:
•
Use lubricants approved for anesthesia or O
2
equipment, such as Krytox.
•
Do not use lubricants that contain oil or grease; they
bum or explode in high O
2
concentrations.
Obey infection control and safety procedures. Used
equipment may contain blood and body fluids.
A moveable part or a removable component may present
a pinch or a crush hazard. Use care when moving or
replacing system parts and components.
When servicing the ventilator, extreme care must be
taken to avoid introducing foreign debris, particularly
metal chips generated by screw threads, into the
pneumatic flow passages of the ventilator. Failure to do
so can result in damage to the flow valve and possible
injury to the patient.
Some internal parts have sharp edges and can cause
cuts or abrasions. Use care when servicing internal
components.
After repairs are completed, always perform the checkout
procedure before returning the system to clinical use.
Refer to Section 3 of this manual.
M1207027
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default
Reproduced from the electronic master in MATRIX