Section 3-13 - Network Configuration
2.) Select the
Configuration
page, then double-click
InSite Exc Agent Configuration(A)
.
NOTE:
This example shows fictional numbers!
3.) Fill out at least all bold stated mandatory fields in the Agent Configuration section(B).
-
The Device Name9Pre-populated) and the CRM field have to be prefix KP.
-
Select Continent and Country from the pull-down menu.
-
Enter City, State, Postal Code and Institution where the system resides.
4.) Check settings in the Advanced Configuration section (C).
-
Enterprise Server - PRODUCT
-
Service Center - EURO
-
Log Level - WARN
Further fields should be pre-populated and should not be modified. However, please ensure correct
setting.
-
Enterprise Server URL - pre-populated URL.
-
Enterprise Tunnel URL -pre-populated URL.
-
File Repository - This path is set by engineering and must not be changed!
-
Dir-D:\InsiteUploadFolder
-
Filter-ensure that this field has the entry “*.txt”s
-
File Watcher: Should always be Enabled!
Figure 3-91 Common Service Desktop - Configuration