Page 35
March 2022
11
EMC Information
Guidance and manufacturer’s declaration –
electromagnetic emissions:
The Wrist Array MR coil is intended for use in the electromagnetic environment specified below. The
customer or the user of the coil should ensure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The Wrist Array Coil uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class A
The Wrist Array Coil is suitable in all
establishments other than domestic, and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Not applicable
Voltage
fluctuation/flicker
emissions IEC 61000-3-3
Not applicable
Not applicable
It is mandatory for the RF shielding effectiveness and filter attenuation in the shielded location to be verified
and to ensure that the actual values meet or exceed the specified minimum values.
Cables or options other than those used for the current product must not be used. If cables or options other
than those used for the current product are used, the EMI performance may deteriorate
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be of wood,
concrete, or ceramic tile. If
floors are covered with synthetic
material, the relative humidity
should be at least 30%
Electrical fast
transient/burst IEC
61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Not applicable
Not applicable