Page 27
March 2022
7
Installation and Maintenance
Installation and configuration
Coil installation and configuration requires the services of your GE Service Representative. Please refer to
the service methods DVD.
For software revision 14.0 and later, the system will automatically identify the coil when connected.
Quality assurance
Quality Assurance testing should be performed upon receipt of the coil to establish a baseline of coil
performance. The testing should then be repeated at regular intervals.
Ask your GE Service Representative how to perform the Quality Assurance on your software level.
7.2.1
Tools required
Description
GE Part #
Dunlee Part #
Quantity
Cubicle Unified Phantom
5342681
N/A
1
Or 1.5T HD 8ch Wrist Array Phantom (only
included with 4535-302-1916*/ 3000-072-9275*)
5160986-5
4535-302-14231
1
Ambient operating conditions
Storage
7.4.1
Storage Location
Symbol
Measurement Type
Operating Conditions
Humidity
30 to 60% non-condensing
Temperature
15-21
O
C (59 to 69.8
O
F)
Symbol
Measurement Type
Operating Conditions
Barometric pressure
500 -1060hPa
Humidity
10 to 100% non-condensing
Temperature
5-65
O
C