Chapter A7: Set Preferences
Definium AMX 700 X-Ray System
5161515-1EN
Rev. 6 (10 February 2008)
A7-35
Operator Manual
© 2008 General Electric Company. All rights reserved.
12. When finished adjusting parameters, press [SAVE] or [SAVE TO MULTIPLE].
[SAVE] applies the parameters to the currently selected patient size.
– A message appears: “Save changes?”
13. Press [YES].
[SAVE TO MULTIPLE] applies the parameters to multiple patient sizes within the selected anatomy.
– The Save to Multiple screen appears.
14. Select the
Patient Size
to save the new look to.
15. Press [SAVE].
16. Press [CLOSE] on the Image Processing Preference Editor.
Noise
Reduction
Suppresses the mottle noise in denser areas of the anatomy while preserving detail
in the rest of the image.
Available options are:
• None – no noise reduction
• Low
• Medium
• High – maximum noise reduction
NOTE:
The noise reduction feature suppresses the mottle noise in denser areas of
the anatomy while preserving detail in the rest of the image. The algorithm
takes into account tissue penetration and dose reaching the detector. For
example, if two Chest PA images were acquired on the same patient, one with
much higher dose than the other, noise reduction may only affect the lower
dose (higher noise) image. In general, the lowest Noise Reduction setting that
produces the desired image quality should be selected.
TE Settings
See Tissue Equalization section for range and effect of TE sliders.
NOTE:
The combined total of the TE Under-penetrated and TE Over-penetrated
areas cannot exceed 100%. The slider will automatically stop and a message
appears: “The Total Area cannot exceed 100%. Reduce the (Under- or Over-)
Penetrated area to proceed.”