Acoustic Output
1
-
90
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
(LOGIQ 7 Only)
NOTE:
a.
This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3)
b.
This probe is not intended for transcranial or neonatal cephalic uses.
c.
This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
Table 1-55: Transducer Mode: M3S Operating Mode: PWD-Mode (inch.B-mode)
Index Label
MI
TIS
TIB
TIC
scan
non-scan
non-
scan
Aaprt<
Aaprt>1
Global Maximum: Index Value
1.7
<1
-
<1
3.3
1.5
IEC
FDA
Units
p
ra
p
r.3
(MPa)
2.7
P
W
o
(mW)
#
-
94.7
94.7
min of [
P
α
(
z
s
,
I
ta,
α
(
z
s
)] [(W
.3(Z1),
I
TA.3
(z1)])
#
z
s
z
1
(cm)
#
z
bp
z
bp
(cm)
#
z
b
z
sp
(cm)
5.1
z
at max. I
pi,
α
z
sp
(cm)
5.0
d
eq
(
z
b
) d
eq
(z
sp
) (cm)
0.5
ƒ
awf
f
c
(MHz)
2.3
#
-
#
2.5
2.5
Dim of A
aprt
X
(cm)
#
-
#
1.5
1.5
Y
(cm)
#
-
#
1.3
1.3
t
d
PD
(µsec)
0.8
prr
PRF (Hz)
635.2
p
r
at max.
I
pi
p
r
@PII
max
(MPa)
3.8
d
eq
at max.
I
pi
d
eq
@PII
max
(cm)
0.5
Focal Length
FL
X
(cm)
#
-
#
6.1
FL
Y
(cm)
#
-
#
5.8
I
pa,
α
at max. MI
I
PA.3
@MI
max
(W/cm
2
)
374.3
Frequency (MHz)
D2.5
#
-
#
D2.5
D2.5
Image Depth (cm)
7.4
#
-
#
6.1
6.1
Vel Scale (kHz)
0.64
-
-
#
1.00
1.00
SV
1
-
-
#
6
6
Penet
Off
-
-
#
Off
Off
A
ssoc
Acousti
c Paramete
Other
i
n
fr
omati
o
n
Operating Cont
rol
Conditions