Acoustic Output Data
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
1
-
59
Direction 5116269-100 Rev. 2
NOTE:
a.
This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3)
b.
This probe is not intended for transcranial or neonatal cephalic uses.
c.
This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
Table 1-24: Transducer Mode: M12L Operating Mode: PWD-Mode (inch.B-mode)
Index Label
MI
TIS
TIB
TIC
scan
non-scan
non-
scan
Aaprt<
Aaprt>1
Global Maximum: Index Value
1.5
2.9
<1
-
1.6
2.3
IEC
FDA
Units
p
ra
p
r.3
(MPa)
3.4
P
W
o
(mW)
92.1
#
16.5
29.7
min of [
P
α
(
z
s
,
I
ta,
α
(
z
s
)] [(W
.3(Z1),
I
TA.3
(z1)])
-
z
s
z
1
(cm)
-
z
bp
z
bp
(cm)
-
z
b
z
sp
(cm)
1.5
z
at max. I
pi,
α
z
sp
(cm)
2.5
d
eq
(
z
b
) d
eq
(z
sp
) (cm)
0.2
ƒ
awf
f
c
(MHz)
5.0
6.5
#
-
5.0
6.5
Dim of A
aprt
X
(cm)
1.8
#
-
0.5
0.6
Y
(cm)
0.6
#
-
0.2
0.2
t
d
PD
(µsec)
1.1
prr
PRF (Hz)
6192.0
p
r
at max.
I
pi
p
r
@PII
max
(MPa)
5.3
d
eq
at max.
I
pi
d
eq
@PII
max
(cm)
0.2
Focal Length
FL
X
(cm)
2.4
#
-
0.8
FL
Y
(cm)
2.1
#
-
1.5
I
pa,
α
at max. MI
I
PA.3
@MI
max
(W/cm
2
)
370.3
Frequency (MHz)
D5.0
BF6.0
#
-
D5.0
D5.0
Image Depth (cm)
1.8
3.0
#
-
1.8
1.8
Vel Scale (kHz)
0.70
-
#
-
3.90
3.90
SV
1
-
#
-
2
2
Penet
On
-
#
-
On
On
As
soc Ac
oustic
Paramete
Ot
her
inf
romati
o
n
Operat
ing C
ontrol
Con
d
itions