Acoustic Output Data
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
1
-
33
Direction 5116269-100 Rev. 2
NOTE:
a.
This index is not required for this operation mode; see section 4.1.3.1 of the
Output Display Standard(NEMA UD-3)
b.
This probe is not intended for transcranial or neonatal cephalic uses.
c.
This formulation for TIS is less than that for an alternate formulation in this
mode.
# No data are reported for this operating condition since the global maximum
index value is not reported for the reason listed.
Table 1-31: Transducer Mode: E8C Operating Mode: PWD-Mode (inch.B-mode)
Index Label
MI
TIS
TIB
TIC
scan
non-scan
non-
scan
Aaprt<
Aaprt>1
Global Maximum: Index Value
1.5
<1
<1
-
2.4
1.1
IEC
FDA
Units
p
ra
p
r.3
(MPa)
3.0
P
W
o
(mW)
#
#
32.0
34.1
min of [
P
α
(
z
s
,
I
ta,
α
(
z
s
)] [(W
.3(Z1),
I
TA.3
(z1)])
-
z
s
z
1
(cm)
-
z
bp
z
bp
(cm)
-
z
b
z
sp
(cm)
1.8
z
at max. I
pi,
α
z
sp
(cm)
1.5
d
eq
(
z
b
) d
eq
(z
sp
) (cm)
0.3
ƒ
awf
f
c
(MHz)
4.1
#
#
-
5.0
5.0
Dim of A
aprt
X
(cm)
#
#
-
1.3
1.3
Y
(cm)
#
#
-
0.5
0.5
t
d
PD
(µsec)
0.5
prr
PRF (Hz)
5379.2
p
r
at max.
I
pi
p
r
@PII
max
(MPa)
3.7
d
eq
at max.
I
pi
d
eq
@PII
max
(cm)
0.3
Focal Length
FL
X
(cm)
#
#
-
2.4
FL
Y
(cm)
#
#
-
2.7
I
pa,
α
at max. MI
I
PA.3
@MI
max
(W/cm
2
)
469.8
Frequency (MHz)
D5.0
#
#
-
D5.0
D5.0
Image Depth (cm)
2.4
#
#
-
2.4
2.4
Vel Scale (kHz)
0.64
-
#
-
1.23
1.23
SV
1
-
#
-
2
2
Penet
Off
-
#
-
Off
Off
A
ssoc
Acousti
c Paramete
Other
i
n
fr
omati
o
n
Operating Cont
rol
Conditions