Quick Start 26
LOGIQ
t
400 Quick Start Guide
2237879–100 Rev. 0
M-Mode Controls (cont’d)
Since M-Mode is basically a single B-Mode scan vector displayed over time, basic controls that affect the B-Mode display
also affect the M-Mode display. See
Quick Start 12 for comments on these controls.
TGC and Depth affect both the M-Mode and B-Mode displays.
If the scan area size is reduced and the position changed, the M-Mode cursor will follow the position change to stay within
the displayed scan area.
The Dual Format keys work the same as in dual B-Mode, but both B-Mode and M-Mode are displayed on the left and right
sides of the screen.
The keyboard controls that effect M-Mode are as follows:
1.
B/M Gain—Controls the gain of the displayed timeline
echoes. To increase/decrease, turn the control clockwise/
counterclockwise.
2.
M/D Cursor—Assigns trackball control to the M-Mode
cursor.
3.
Zoom—Enables real-time or freeze zoom function.
M-Mode Optimization
Adjustments for...
Do the Following...
Adjustments for...
Do the Following...
Improve M-Mode
1.
Increase/decrease B/M GAIN.
2.
Increase Dynamic Range.
Increase Size of Area
of Interest
1.
Use M-Mode ZOOM.
1
3
2