Troubleshooting
For more information on setting up LAN communication, see
Chapter “Options
2.
Verify the ethernet cable is connected properly to the ethernet port of the Silex
Wireless Bridge and the other end of the ethernet cable to ethernet port of
the MAC2000 device.
For information on where the Silex Wireless Bridge connects to the device, see
Chapter “Rear View” on page 32
3.
Check communication setup to verify whether the
IP
,
netmask
,
gateway
, and
DNS
server addresses are all correct.
For details on checking addresses, see
“Communication Setup” on page 140
Cannot Transmit Records via USB Wireless
If you receive an error while attempting to transmit records via USB wireless, use
the following procedure:
1.
Verify the correct communication option was activated.
The system supports two options for communicating over wireless using USB
wireless dongle:
• WIFC (for communicating with a CardioSoft/CS system
• WIFM (for communicating with a MUSE system.
For more information on setting up LAN communication, see
on page 165
.
2.
Verify the wireless dongle is connected properly to the USB port on the rear
of the device.
If the wireless dongle is not working, try changing the port.
For information on where the wireless dongle connects to the device, see
.
3.
Check communication setup to verify whether the
IP
,
netmask
,
gateway
, and
DNS
server addresses are all correct.
For details on checking addresses, see
“Communication Setup” on page 140
4.
Check whether the wireless LAN is enabled and the authentication details are
correct.
For information on the wireless LAN setup, see “Wireless Networking Settings” in
“Communication Setup” on page 140
Cannot Transmit Records via Embedded Wireless Module
If you receive an error while attempting to transmit records via the embedded wireless
module, use the following procedure:
1.
Verify the correct communication option was activated.
The system supports two options for communicating over a wireless network
using the embedded wireless module:
• WIFC for communicating with a CardioSoft/CS system
• WIFM for communicating with a MUSE system
188
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......