CU-960 only:
Installation on ÄKTAready
Connection from the system computer to the internal control unit CU-960 is made
by the water proof RJ45 connector placed in the bottom of the main cabinet
(VS-08-A-RJ45/MOD-1-IP67 + VS-08-BU-RJ45-5-F/BU). It is absolutely necessary
from an EMC perspective to connect this connector shielding to the cabinet using
either copper tape or the small sheet-metal shroud developed in the project.
The mating connector cable assembly needs to have the matching water proof
plug type provided by Phoenix. (VS-08-RJ45-5-Q/IP67) to maintain IP67. Suitable
mating cable assembly is described in assembly drawing 14001358 (part number
28-4022-13) and is supplied with the system in the accessory-kit.
Note:
The internal LAN cable is a cross-over type!
If the system is connected to a switch/hub, a cross-over cable must be connected
in series with a cable provided by GE Healthcare, part number 28-4022-13.