Instructions - 28-9414-37 Edition AA
Control Units CU-950 and CU-960
How to connect the cables:
1
Connect the UniNet-1 data communication cable between the UniNet-1 port
on CU-950 and Pump P-920 in the system.
Note:
If a fraction collector is used, connect the fraction collector UniNet-1
cable between P-920 and the fraction collector, and the 1.5 m UniNet-
1 cable between the fraction collector and CU-950, and then the USB
cable between CU-950 and the computer.
2
Connect the cable on the CU-950 power converter to the Input port on
CU-950.
3
Connect the mains cable between the power converter and the mains supply
socket at the rear of the system.
4
If an autosampler is used, the mains cable to the autosampler may be
connected to an external mains outlet if all mains supply sockets on the
system are occupied.
Note:
In this case the autosampler must be manually reset by disconnecting
the power cable when performing a system reset or restart.
5
Connect the USB cable to the USB port on CU-950. For connection to and
installation on the computer, please refer to the UNICORN Administration and
technical manual.