Instructions - 28-9414-37 Edition AA

 

Control Units CU-950 and CU-960

1100 

Error in strategy 
execution 

Internal error.

Create a report.

1101 

Undef. strat block, 

The strategy 
contains an 
unknown block 

Upgrade UNICORN software 
or reinstall old strategy

1700 

>95% CPU Load 

Internal error.

Create a report.

# SAP error codes

1303 

Tuning forbidden in 
this state 

Internal error.

Create a report.

1305 

Wrong test type in 
strat. block 

Internal error.

Create a report.

1308 Wrong 

parameter 

type in strat. 

Internal error.

Create a report.

1309 

Parvalue out of range 
in strat 

Internal error.

Create a report.

1310 

Missing par in 
strategy block 

Internal error.

Create a report.

1311 

Wrong index in strat. 
RUNMODE 

Internal error.

Create a report.

1312 Strategy 

contains 

unknown block 

Internal error.

Create a report.

1315 

Trm index out of 
range, RUNMODE

Internal error.

Create a report.

2000 UniNet 

controller 

is 

restarted 

The UniNet 
controller is 
restarted - 
messages are lost.

Check instrument connection. 

2001 

WARNING - faulty 
UniNet comm. 

Faulty UniNet 
communication- 
no message lost.

Check instrument connection. 

2002 UniNet 

controller 

overrun 

Messages are lost. Check instrument connection. 

2003 

UniNet pckg rec in 
wrong order 

UniNet package 
received in wrong 
order. 

Make sure that all connected 
valves have a unique id. 
Number.

2010 

Too many pars

Internal error.

Create a report.

2011 

Undef service 

Internal error.

Create a report.

2012 

No response 

System could not 
find specific unit. 

Check instrument connection.

If Error message appears 
during system start up and 
specific unit is not included in 
the system configuration this 
message can be ignored.

Error Id no. Error message 

Cause

Suggested Action

Summary of Contents for CU-960

Page 1: ...GE Healthcare Control Units CU 950 and CU 960 Instructions...

Page 2: ......

Page 3: ...est The CE symbol and corresponding Declaration of Conformity is valid for the instrument when it is used as a stand alone unit or connected to other CE marked GE Healthcare instruments or connected t...

Page 4: ......

Page 5: ...el 8 3 CU 950 only Installation on KTAexplorer and KTApurifier 9 4 CU 950 only Installation on KTAFPLC 11 5 CU 950 only Installation on KTApilot 13 6 CU 960 only Installation on KTAprocess 14 7 Flash...

Page 6: ...60 are used as the controlling interface between a UNICORN workstation and the components of an KTA system CU 950 This instruction describes CU 950 and provides information on how to install this cont...

Page 7: ...systems The CU 950 Advanced mode is offered with an option to use 128 MB internal memory for improved preparedness in case of communication failures between the UNICORN workstation computer and the C...

Page 8: ...changes from Scanning to Ready 2 2 Back panel The back panel ports are from left to right Power supply input CU 960 only Profibus port for system communication USB port for network controlling worksta...

Page 9: ...ration below KTApurifier system is shown as an example Figure 3 1 Location of CU 950 on KTApurifier system Figures 3 2 and 3 3 show the schematic CU 950 cable connections for KTAexplorer and KTApurifi...

Page 10: ...ort on CU 950 3 Connect the mains cable between the power converter and the mains supply socket at the rear of the system 4 If an auto sampler is used the mains cable to the autosampler may be connect...

Page 11: ...side of UPC 900 Figure 4 2 Schematic UniNet 1 and mains cable connections for KTAFPLC Power converter Mains supply socket Mains cable Voltage WARNING For continued protection against risk of fire repl...

Page 12: ...mputer 2 Connect the cable on the CU 950 power converter to the Input port on CU 950 3 Connect the mains cable between the power converter and the mains supply socket at the rear of the system 4 If an...

Page 13: ...fraction collector UniNet 1 cable between the controller rear panel inside KTApilot and the fraction collector and the 1 5 m UniNet 1 cable between the fraction collector and CU 950 2 Connect the cabl...

Page 14: ...very The unit is mounted in a rack adaptor which is mounted in the instrument rack inside the electronics cabinet of the KTAprocess system as illustrated below Figure 6 1 CU 960 mounted in an KTAproce...

Page 15: ...til communication between the workstation and the CU 960 is re established or the workstation is functional again The run will continue only until the flash memory is full Run data is saved at interva...

Page 16: ...e a report 1008 No decimal point in real Internal error Create a report 1009 Value table index out of range Internal error Create a report 1010 No V when expected Internal error Create a report 1011 N...

Page 17: ...ngmethodevent list Internal error Create a report 1040 Erroneous condition id Internal error Create a report 1041 Missing submethod list Internal error Create a report 1042 Too many parameters in inst...

Page 18: ...egy contains unknown block Internal error Create a report 1315 Trm index out of range RUNMODE Internal error Create a report 2000 UniNet controller is restarted The UniNet controller is restarted mess...

Page 19: ...reate a report 38 See Help Internal error Create a report 51 Unable to allocate CAN c block Internal error Create a report 52 Unable to post CAN cmd block Internal error Create a report 53 Unable to a...

Page 20: ...ing to the cabinet using either copper tape or the small sheet metal shroud developed in the project The mating connector cable assembly needs to have the matching water proof plug type provided by Ph...

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ...l third party trademarks are the property of their respective owners 2009 General Electric Company All rights reserved First published Jan 2009 All goods and services are sold subject to the terms and...

Reviews: