GDS-68XP Operation & Maintenance Manual, Revision 3.5
Page 49
Ch
2
Balance
43295
3
N/A
Binary
Ch
2
Heater
Enable
43296
3
N/A
False
=
NO,
True
=
YES
Ch
2
Heater
Setpoint
43297
3
N/A
Modbus
32
bit
IEEE
754
Floating
Pt
Ch
2
Temp
Comp
‐
40
43299
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Temp
Comp
‐
30
43303
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Temp
Comp
‐
20
43307
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Temp
Comp
‐
10
43311
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Temp
Comp
0
43315
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Temp
Comp
+10
43319
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Temp
Comp
+20
43323
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Temp
Comp
+30
43327
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Temp
Comp
+40
43331
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Temp
Comp
+50
43335
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Temp
Comp
+60
43339
3
N/A
32
bit
FP
Gain,
32
bit
FP
Offset
Ch
2
Sensor
Type
43343
3
N/A
TBD
Ch
2
Send
Sensor
Life
43344
3
N/A
False
=
NO,
True
=
YES
Ch
2
Cal
mA
Setting
43345
3
N/A
Modbus
32
bit
IEEE
754
Floating
Pt
Ch
2
Local
Cal
43347
3
N/A
False
=
NO,
True
=
YES
Ch
2
AI
Range
43348
3
N/A
TBD
Registers
45001
‐
45022
added
in
V1.03
Ch
1
Value
45001
3
N/A
800
=
“0”,
4000
=
Full
Scale
Ch
2
Value
45002
3
N/A
800
=
“0”,
4000
=
Full
Scale
Ch
1
Value
45003/04
3
N/A
MODBUS
32
bit
floating
point
Ch
2
Value
45005/06
3
N/A
MODBUS
32
bit
floating
point
Ch
1
Alarm
1
Status
45007
3
N/A
“1”
=
Fault
Ch
1
Alarm
2
Status
45008
3
N/A
“1”
=
Fault
Ch
1
Alarm
3
Status
45009
3
N/A
“1”
=
Fault
Ch
1
Fault
Status
45010
3
N/A
“1”
=
Fault
Ch
2
Alarm
1
Status
45011
3
N/A
“1”
=
Fault
Ch
2
Alarm
2
Status
45012
3
N/A
“1”
=
Fault
Ch
2
Alarm
3
Status
45013
3
N/A
“1”
=
Fault
Ch
2
Fault
Status
45014
3
N/A
“1”
=
Fault
Ch
1
Sensor
Life
45015
3
N/A
0
‐
100
binary
Ch
2
Sensor
Life
45016
3
N/A
0
‐
100
binary
Ch
1
Sensor
Temp
45017
3
N/A
Binary
0
‐
4095
Ch2
Sensor
Temp
45018
3
N/A
Binary
0
‐
4095
Ch
1
Sensor
Temp
45019/20
3
N/A
MODBUS
32
bit
floating
point
Ch
2
Sensor
Temp
45021/22
3
N/A
MODBUS
32
bit
floating
point