35
T E C H N O L O G I E S
Input
24v DC ±20% 170mA
Flow Rate
2.5 litres/min
Continuous
Pump Motor - 6 years expected
Running
Diaphragm - 3 years in air
Flow Fail
Green LED – normal operation
Monitor
Red LED – pipe blockage (FF)
Red LED – pump fail (PF)
Sounder – not less than 85dB @ 30cm
Outputs
Relay S.P.C.O 1Amp @ 230vAC
N/D- N/E option
Selectable 5/12vDC PF/FF/PF+FF
Factory set – open drain
Sample Line
6mm OD 4mm ID
Max Length – 100m, min – 1m
Material – nylon, polypropylene, PTFE
Couplings
In/Out - push fit - barbed
Fixing
4 – 3.5 Ø
Dimensions
123L 54W 80H
Weight
320 gms
Mounting/Position
The pump sensor unit should be mounted at chest height with
the sample line travelling to the chosen target area at either
maximum height for lighter than air gases or minimum low
level (100mm) for heavier than air gases. A filter (Part no. 008-
212) should be attached to the pipe end where high levels of
particulates may be present.
Field Terminals
Signal Out Jumper Selection
SENSOR ASPIRATOR MODULE
TYPE ASM-2
PUMP/FLOW MONITOR
NC C NO + - FF PF
12v
5v
FF
Relay
24vDC
IN
OUT
NE
FF
5V
MUTE
NC C NO + - FF PF
12v
5v
FF
Relay
24vDC
IN
OUT
NE
FF
5V
MUTE