background image

30

T E C H N O L O G I E S

Fig.1

Fig.2

Fig.3

Technical Sheet

 

ref C1698 Issue Bv.3

The Combi sensors which have an LCD display fitted also 
incorporate 3 reed switches which can be activated using external 
magnets through the glass window of the flameproof XDIwin 
enclosure. 

These magnets do not act instantly and have to 

be in close proximity to L, M and R on the front display for 

a few seconds to activate a software setup function. 

The left magnet 

enters the Auto zero ON or OFF menu.  

This allows small drift changes in the sensor to be compensated 
for but is not operational when the sensor readings are greater 
than 5% of full scale. Therefore auto zero is inactive when a larger 
gas reading is present. When the remove magnets message 
appears, move the left magnet away and then the display shows 
if auto zero is ON or OFF. The left magnet puts auto zero ON 
and the right magnet turns it OFF. With no magnets present, the 
display will return to normal after a few seconds timeout. 

The right magnet

 allows the CAN address of the sensor to be 

changed. When the ADDRESS menu is displayed with a prompt 
to remove the magnet, and then the display shows the address 
and that the right magnet decreases it whilst the left magnet will 
increase it. This is then stored in internal non volatile memory 
and the display will automatically revert to normal operation.     

The centre magnet 

is used to inhibit the sensor. As with the left 

and right magnet functions the display requests that you remove 
the magnet and then the state of the inhibit appears on the LCD. 
The left magnet then puts the sensor into inhibit whilst the right 
magnet removes it. An amber LED on the front panel under the 
LCD flashes when the sensor is inhibited. When all magnets are 
removed, the display will revert to normal operation.

The left and right magnets together allow the calibration 

menu to be used.

Removing both magnets as instructed on the LCD presents the 
first part of this multi menu which is ZERO. With no gas present 
use the left magnet to increase the reading and the right magnet 
to decrease to achieve a zero reading on the display. A timer 
is displayed on the LCD and when this reaches 0, the next 
menu is displayed. This timer is 15 seconds approximately and 
is reset back each time a magnet is near. Waiting till timeout 
is acceptable but this timeout can be speeded up by placing a 
magnet near to the centre position.

SPAN is the next part of the menu and gas should be applied to 
the sensor at this time.

The left magnet increases the gain and the right magnet reduces 
gain. The actual sensor value can be seen on the display to rise 
or fall respectively.

LOW ALARM is the next menu and left and right magnets increase 
and decrease this value.

HIGH ALARM is next followed by OVER RANGE alarm.

The direction of the alarms is displayed as ^ for rising and v for 
falling but these can be changed using left and right magnets 
together.

EOL

CAN

4-20

MPU

FAULT

TP1

OV

TP11

TP9

TP5

TP8

TP10

198M1C

ISS M

TP

7

ZERO

ZERO

TP12

DISP

MPU active LED

(Flashing)

Fault LED

Sens V Test Point

TP7 + TP12

Sensor cell input

Sensor Volts

(RV3)

Zero

(RV2)

Span

(RV1)

4mA

(RV4)

Sensor Zeroing LEDs

CAN Transmitting LED

4-20mA  SInk/Source

4-20mA  Test

Jumper

End-of-line Jumper

Relay/Win Ribbon

LCD Contrast

Win Only

RS232

PC Data In/Out

Win Data Ribbon Cable

4-20op  Test Points

FIELD TERMINALS

HI      LO

24V Can+ Can- OV 4-20mA

24V Hi

Lo OV 4-20

mA

P

Y

W

Control Unit

CAN 1 or 2

Sensor

Addressable

+24  
Hi
Lo
0v
4~20mA

+24

Hi

Lo

0v

Control Unit

4~20mA Input

Sensor

3-Wire Direct

4~20mA Signal

+24  
Hi
Lo
0v
4~20mA

+24

Sig

0v

EOL

T

FAULT

MPU

SK

SE

204M1C

ISS M

CONTRAST

Fault LED

Gas Type Header

Sensor cell input

20mA

Span

(RV2)

4mA

Zero

(RV1)

Sensor 4-20mA Routing Configurations

(see diagram)

CAN Transmitting LED

4-20mA  Test

Jumper

Relay/Win Ribbon

LCD Contrast

Win Only

End-of-line Jumper

+24V Test Point

(w.r.t 0V)

+24V

RS232

PC Data In/Out

Win Data Ribbon Cable

4-20mA  Test Pins

4-20mA  Sink Source

External 4-20mA  Input

(eg. Infrared)

+24V O/Pt

(eg. Infrared)

FIELD TERMINALS

HI      LO

24V Can+ Can- OV 4-20mA

24V Hi

Lo OV 4-20

mA

P

Y

W

Mainboad to Sensor wiring

Sensor Jumper configuration

Control Unit

CAN 1 or 2

Sensor

Addressable

+24  
Hi
Lo
0v
4~20mA

4~20 IN

SO

SK

+24

Hi

Lo

0v

Control Unit

4~20mA Input

Sensor

3-Wire Direct

4~20mA Signal

+24  
Hi
Lo
0v
4~20mA

+24

Sig

0v

Control Unit

4~20mA Input

Sensor

2-Wire Direct

4~20mA Signal

+24  
Hi
Lo
0v
4~20mA

+24

Sig

0v

Control Unit

CAN 1 or 2

Sensor

Addressable 4~20mA Input

+24  
Hi
Lo
0v
4~20mA

+24

Hi

Lo

0v

4~20 IN

SO

SK

4~20 IN

SO

SK

4~20 IN

SO

SK

ange

t magnets 

J10

J10

J10

Using magnets (set up)

Display 

Contrast

Field terminals

Relay PCB  
Connector

Display PCB  
Connector

Test

Test Pins

24v DC
OV
FLT

OUT
IN

AG

Zero LED

ADDR

4mA

20mA

Sensor

Zero

Sensor Volts

Comms 

Port

EOL

Vo

Summary of Contents for GasVac LG-1

Page 1: ...S TECHNOLOGIES LTD FUSION POINT ASH LANE GARFORTH LEEDS LS25 2GA UK www gds technologies co uk This document is not contractual and the equipment specification may be modified at any time without prio...

Page 2: ...9 LOCAL AND REMOTE RELAYS 9 MODBUS SET UP AND INDICATORS 10 REPEATER 11 RS232 DATA LOG OUTPUT 11 EVENT LOG 12 DATA TO AND FROM A PC USING GDS TERMINAL 12 SETTING THE TIME TO ALARM 14 SETTING ALARM LEV...

Page 3: ...s to the PC Using simple text files these can be edited and sent back to the panel In addition to 8 gasses an optional cabinet flammable only gas is monitored in the panel and will cause power shutdow...

Page 4: ...tional 64 relays Common Low alarm relay DPCO or Group 1 Common High alarm relay DPCO or Group 2 Common Over Range alarm relay SPCO or Group 3 Fault relay SPCO or Group 4 OPT G5 LTEL STEL relay SPCO or...

Page 5: ...specific sensor if it is in alarm Pressing button 4 will stop the cycle on the current sensor for 999 seconds If a new alarm occurs the display will show this sensor The type of gas and the text asso...

Page 6: ...button again Button 4 is the button Pressing this allows the display to hold the automatic sequence of sensors being displayed for 999 seconds When in engineer s mode it is used to increment parameter...

Page 7: ...assword cannot be changed System Menu Index See Appendix 1 for further details Note The items that appear in the user menu can be selected from the Engineer s menu by toggling the 4 button A symbol wi...

Page 8: ...the system back to normal TESTING THE LEDS 7 AND RELAYS 8 To test all LED s and beep the buzzer use menu number 7 and press SELECT 3 to make all LED s illuminate in dimmed mode to see if they are fun...

Page 9: ...il mist B2H6 32 Diborane 0 1ppm SiH4 33 Silane 0 5ppm Ph3b 34 Phosphine 0 1ppm F2 35 Flourine 0 5ppm NF3 36 Nitrogen Triflouride 0 50ppm LEL M 37 LEL per metre 0 5 SW Input 38 Switch input 0 100 Displ...

Page 10: ...essor board will sound By pressing the ACCEPT button 1 on the front panel the buzzer will become silent FLOW FAIL AND PUMP FAIL The pressure of the pump s in the sample system are monitored and a loss...

Page 11: ...baud rate expressed in bits per second is set using menu 34 All communicating devices in a Modbus system should be set to the same baud rate and the system is capable of supporting 19200 9600 and 4800...

Page 12: ...5 buttons to choose a free address that the repeater will use as its own address Note this has to be different from that of the main panel which defaults to address 201 Press SELECT 3 Set FROM ADDRESS...

Page 13: ...ext data file from a PC using the RS232 interface via GDS Terminal which can be downloaded from the GDS website The file which is sent is simple alphanumeric characters which can be edited using a sim...

Page 14: ...ts 10 times the actual value this is to avoid decimal points in the data LoAlarm 1 200 Ditto for sensor 1 low alarm point OrAlarm 3 998 Ditto for sensor 3 over range alarm X This terminates the text t...

Page 15: ...ent on sensors Should a sensor cell require to be replaced the sensor should have its power removed before changing the cell itself Flammable Catalytic and Toxic cells have colour coded wiring which i...

Page 16: ...n the main PCB will be activated according to the voting set up and de activated when the condition has subsided and after the relay off timer has expired Using UP 1 and DOWN 5 to select the group the...

Page 17: ...ing group becomes a TWA Time Weighted Average group and the normal voting function is suppressed DEFAULTING THE PANEL Individual sections of the panel can be defaulted rather than restoring all the de...

Page 18: ...area only Positioning of the instruments should be chosen with regard to the following points Away from sources of local heat with room for adequate ventilation Within easy reach and audible distance...

Page 19: ...the catch pot for high levels of particulate or excessive water contamination On sites involving a high risk process of having gases which may cause loss of sensitivity a check on calibration should...

Page 20: ...exit this menu 6 Edit Company Text Edit Fault Input Text The fourth line of the LED display alternates between time and company text This text can be edited to show the site or installer details 7 La...

Page 21: ...LCD indicates whether inhibited or not and the UP 1 and DOWN 5 buttons change the channel Note that only channels with active sensors can be inhibited Pressing SELECT 3 or timeout will move to the sec...

Page 22: ...be redirected to 7 relays on the 8 way relay card The fault relay 4 always remains as Fault but the remaining relays are assigned as remote relays 123 5678 respectively This feature eliminates the nee...

Page 23: ...anged to a new value using the 4 and 2 button but only if an address link has been inserted This link should be removed afterwards Note If more than 1 device has a link inserted then all these will ha...

Page 24: ...from this menu you will be prompted to enter a new code twice via the buttons Pressing SELECT 3 will exit this menu Note The user password cannot be changed 31 Set Modbus Baud This menu allows you to...

Page 25: ...use pump room etc G1 LeL x16 Direct Gas type unlatchedAlarm 1 x16 Direct alarms latchedalarm or unlatchedalarm unlatchedAlarm 1 x64 Addressable alarms latchedalarm or unlatchedalarm LoDirAlarm 1 200 x...

Page 26: ...t relay patterns x64 As above but for CAN Fault relay patterns h01 0000000000000000 h02 0000000000000000 gfedcba987654321 CAN Low Alarm relay patterns x64 As above but for CAN Low Alarm relay patterns...

Page 27: ...han 182 Sensor fail timeout 17 32 channel 1 alarm bit 0 is chan 17 etc Can Chan 183 Sensor fail timeout 33 48 channel 1 alarm bit 0 is chan 33 etc Can Chan 184 Sensor fail timeout 49 64 channel 1 alar...

Page 28: ...y D type plugs into the PC COM port and the pre wired sensor connector is inserted into J3 on the Combi Flam PCB 198 Issue P and J3 on the Combi Tox PCB 204 Issue N When Hyper terminal on the PC set a...

Page 29: ...28 Front Panel PCB...

Page 30: ...esent I shows that this sensor has its alarms inhibited under the O H L the v represent the direction of the alarms L is falling and H and O are rising A under the letter s OHLDFI represents a detecte...

Page 31: ...t to decrease to achieve a zero reading on the display A timer is displayed on the LCD and when this reaches 0 the next menu is displayed This timer is 15 seconds approximately and is reset back each...

Page 32: ...ess A to change the address of this sensor if required d Press N to select the number of decimal places to 1 or 2 ie dp 1 or dp 2 e With no gas applied and 4mV measured at test pins TP8 TP9 press Z to...

Page 33: ...t turns it off With no magnets present the display will return to normal after a few seconds timeout The right magnet allows the CAN address of the sensor to be changed WHEN the ADDRESS menu is displa...

Page 34: ...justed by the DPM Z potentiometer zero CALIBRATION With the digital meter connected to the test pins TP1 and TP2 and a reading of 4mV clean air apply test gas and wait until a maximum reading is obtai...

Page 35: ...4 CAN Address 245 1 2 3 4 CAN Address 249 1 2 3 4 CAN Address 253 1 2 3 4 1 2 3 4 CAN Address 246 1 2 3 4 CAN Address 250 1 2 3 4 CAN Address 254 1 2 3 4 1 2 3 4 CAN Address 247 1 2 3 4 RELAYS 57 64...

Page 36: ...e PTFE Couplings In Out push fit barbed Fixing 4 3 5 Dimensions 123L 54W 80H Weight 320 gms Mounting Position The pump sensor unit should be mounted at chest height with the sample line travelling to...

Page 37: ...P1 CA Flow Oxygen 4 Cab Gas Monitor Flam 5 P1 Flow Block 13 8 way relay PCB 14 5A Power Supply Din Rail 18 0 2 5 L min Flow Gouge 19 Flow Regulator 22 Toggle Switch 31 Water Trap Catch Pot 52 TEE Tube...

Page 38: ...Trap 3 10 0096 1447 R4 S8 25 Combi Panel Shield 1 13 0030 1603 R4 S4 26 Combi 80 Label 1 12 0032 977 R24 S3 27 Bulkhead fitting 2 10 0055 196 R3 S5 28 Wall Mount Bracket 1 11 0015 765 R4 S7 29 M4 Plas...

Page 39: ...0 2 Ltr min Pump Line Blockage and Pump Fail Monitor Test Gas Point 0 5 1 Ltr min Flame Arrester Flame Arrester 24A 23 19 51 24B 24C 5 18 22 52 31 Catch pot 1 2 For vacuum or atmospheric sampling do...

Page 40: ...39...

Page 41: ......

Page 42: ......

Page 43: ......

Page 44: ...hnologies co uk Phone 44 0 113 286 0166 E sales gds technologies co uk T 0113 286 0166 F 0113 287 8178 GDS TECHNOLOGIES LTD FUSION POINT ASH LANE GARFORTH LEEDS LS25 2GA UK www gds technologies co uk...

Reviews: