GASMAX CX + GDS-IR Operation & Maintenance Manual, Revision 1.0
Page 10
LOCAL
GDS
‐
IR
SENSOR
WIRING
A
GASMAX
CX
with
local
GDS
‐
IR
sensor
is
pre
‐
wired
at
the
factory
as
shown
below.
The
GASMAX
CX
includes
a
sensor
personality
board
that
adapts
the
GDS
‐
IR
wiring
to
the
GASMAX
CX
inputs
and
provides
a
magnetic
switch
that
allows
the
user
to
perform
the
“IR
Zero”
function
without
opening
the
enclosure.
Figure 3-4: GASMAX CX + GDS-IR Local Sensor Wiring Diagram
REMOTE
GDS
‐
IR
SENSOR
WIRING
Wiring
between
the
GASMAX
CX
and
a
remote
GDS
‐
IR
sensor
is
provided
by
the
customer.
The
interface
requires
a
four
‐
conductor
shielded
cable
as
shown
below.
The
remote
junction
box
provides
both
a
magnetic
switch
and
local
pushbutton
for
the
“IR
Zero”
function.
Figure 3-5: GASMAX CX + GDS-IR Remote Sensor Wiring Diagram
CO
M
420
OU
T
2
420
OU
T
1
PW
R
R/
P
W
R
C/
4
20
IN
1
A
/CO
M
42
0
IN
2
CH
1
CH
2
E
‐
NE
T
LOCAL
SENSOR
TO DISPLAY
Black
Wire
–
Common
Ground
Blue
Wire
–
Sensor
4
‐
20mA
Output
White
Wire
‐
IR
Zero
Input
Red
Wire
‐
+18
to
+30VDC
Input
Important
–
Remove
DC
power
before
connecting
or
disconnecting
any
wiring
to
/
from
the
GDS
‐
IR
Sensor
Personality
Board
Sensor
Personality
Board
Remote
GDS
‐
IR
Sensor
Remote
Sensor
Junction
Box