80021997 Ver. B, 2017-04
DU-WA-0016-01 Rev B 4/18/17 GCX Corp Page 1 of 4
Tighten Knob
3875 Cypress Drive, Petaluma, CA 94954 800.228.2555 707.773.1100 Fax 707.773.1180 www.gcx.com
GCX Mounting Assembly Installation Guide
M-Series Wall Mount Kit for Welch Allyn® Connex® VSM 6000 Series
and VSM 300 Series Monitors
The purpose of this guide is to describe installation of the Wall Mount Kit.
Parts Reference
The parts list below includes parts and hardware that will be used in this installation procedure (hardware not shown).
Tools Required:
Phillips screwdriver (not provided), 1/2 in. socket wrench (not provided), 1/8 in. hex wrench (provided).
Installing the Wall Mount Kit
1. Install the Wall Channel in accordance with the
GCX Wall Channel Kit Installation Instructions (DU-WC-0002).
2. Insert one (1) Adjustable Stop in the top of Channel and let it slide to rest against the Fixed Stop at bottom of the
Channel (left photo below).
3. Install the Utility Basket by inserting the Slide (on the back of the Basket) into the top of the Channel and sliding the
Basket to the desired mounting position (center photo below). Using the 1/8 in. hex wrench, tighten the Set Screw to
secure the Basket in position. Slide the Adjustable Stop up the Channel to the bottom of Basket Slide and tighten the
center screw in the Adjustable Stop.
4. Insert the remaining Adjustable Stop in the Channel and let it slide to the top of the Basket Slide (right photo below).
Adjustable Stop #2
Adjustable Stop
Adjustable Stop #1
Tighten Set Screw
Item # Description
Quantity
VSM300
Series
VSM6000
Series
1
Wall Channel Kit, 19” (includes instructions – not shown)
1
2
M-
Series Arm with Mounting Adapter, 8”
1
3
Utility Basket
1
4
Adjustable Stop
2
5
18” Strap, Hook and Loop (not shown)
2
0
6
1/8” Hex Wrench (not shown)
1
7
Wall Channel Cover, 16” (includes instructions – not shown)
1