background image

DU-AG-0021-100 Rev A 4/15/08 GCX Corp. Page 2 of 6 

Attaching Top Plate/Channel to Top of Fabius GS 

The following parts and hardware are included in this installation kit (hardware not shown):

 

 

 

 

 

 

 

Tools Required 

•  Drill (not provided) 
•  1/4'' drill bit (not provided) 
•  Phillips screwdriver 

 

1.  Locate four (4) dimples in top of anesthesia machine. Dimples indicate locations of threaded inserts beneath top 

surface. If dimples are difficult to locate, use Top Plate/Channel assembly as a template and mark top of machine 
through four (4) mounting holes. 

 

2.  Using a 1/4'' drill bit, drill in locations marked in step 1 (drill only where dimples are present). 

CAUTION:

 Do not allow 

drill bit to penetrate more than 1/4'' into top of machine. Drilling too deeply may damage threaded inserts and internal 
components of the machine. Remove all drilling chips. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
3.  Align four (4) mounting holes in Top Plate with corresponding holes in top of machine (drilled in step 2). Fasten Top 

Plate to machine with four (4) M5 x 30mm Pan Head Machine Screws (PHMS)  

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Item # 

Description 

Qty 

Top Plate with 21'' Channel 

M5 x 30 mm Pan Head Machine Screw (PHMS) 

Label, Tip Hazard 

Top Plate/Channel

M5 x 30mm PHMS (4) 

Locations of Threaded Inserts (4) 
in top of Fabius GS