Summary of Contents for ECOlite4000

Page 1: ...EN Page 1 12 ECOlite 4000 GAS SAVER INSTRUCTION FOR USE...

Page 2: ...om the oxygen The dosage and timing of oxygen delivery is controlled by a microprocessor in accordance with the chosen setting Dosage per breath is independent of breathing rate therefore additional o...

Page 3: ...TIONS At atmospheric pressure up to 60 oxygen is harmless 100 oxygen produces a cough after 6 hours and increasing breathlessness after 25 hours PERSONNEL INSTRUCTIONS AND TRAINING According to Medica...

Page 4: ...function from national standards In the event of any problem with connection please contact the distributor the manufacturer Display screen ON OFF button Oxygen inlet Outlet into cannula Control butt...

Page 5: ...clean If the unit is being used with an oxygen supply source with adjustable fl ow rates set the flow rate on the supply source to 4 litres per minute or to the closest setting above Check the cylinde...

Page 6: ...s the button and hold it down Afterwards press the ON OFF button and also hold it down A PRESS KEY sign will appear Press the two buttons marked and simultaneously a LANGUAGE sign will appear Using th...

Page 7: ...ngly Activate the alarm by not breathing in the cannula After approx 1 minute the alarm should set off both by sound and an alert on the display If there is an abnormal sound noise gas leakage or any...

Page 8: ...void wasting oxygen 6 4 2 DISCONNECTING THE DEVICE Before disconnecting the ECOlite 4000 from the gas source Close the oxygen supply from the cylinder reservoir disconnect the connector of the spiral...

Page 9: ...At the end of the product s life time 10 years at most the ECOlite 4000 device must be withdrawn from service To prevent reuse the owner shall make the ECOlite 4000 impossible to use or return it to...

Page 10: ...c Area EEA This mark on the product indicates compliance with Directive 93 42 EEC relating to medical devices Class IIa Functional performance Sensitivity 0 13 cm of H2O Setting Manual automatic Trigg...

Page 11: ...eter Use by Outlet parameter Take back equipment for recycling Do not dispose equipment into unsorted municipal waste WARRANTY The Standard Warranty period is two years from date of receipt by the GCE...

Page 12: ...Asia The company operates 15 subsidiaries around the world and employs more than 850 people GCE Group includes four business areas Cutting Welding Process Applications Medical and High Purity Today s...