CLEANING

Cleaning of the device

Clean the device as necessary with a clean dry cloth. Do not immerse it in water or any other 

cleaning liquid. The nasal cannula is designed for single patient use.

Do not immerse the device in water or any liquid.
Do not use rough or abrasive cleaning materials as damage may result.
Do not expose the device to high temperature (such as to an autoclave).
Do not immerse the cannula or spiral hose in a liquid disinfectant; the user cannot be 

certain they have eliminated the remaining moisture inside the hose, which might damage 

the unit, possibly endanger the patient.

MAINTENANCE

9.1 MAINTENANCE

Prior to each use:

• 

Check the connections to the unit. 

• 

Make sure that the battery does not require replacing. 

• 

Check the device for physical damage. 

• 

Conduct a functional test (see Chapter 6.3.5). 

If any problem is identified with the unit, contact the supplier of the ECOlite

®

 4000.

After every period of use it is recommended to:

• 

Change single-use components (nasal cannula) 

• 

Clean the device as described above.

9.2 LIFETIME AND SERVICE INTERVAL

The maximum lifetime of the product is 10 years. At the end of the product’s life time (10 years at 

most), the ECOlite

®

 4000 device must be withdrawn from service. To prevent reuse, the owner shall 

make the ECOlite

®

 4000 impossible to use, or return it to GCE. Contact your local representative 

for further details before returning the product.

No service check is needed during the 10 years lifetime of the ECOlite

®

4000, however all 

repairs to the product must be done by GCE authorised personnel.

GCE coding data

The serial number is made up as follows: YY MM XXXXX A

Y: 2 last numerals of the year

M. month X: serial number

A: manufactured at GCE s. r. o. CZ

The owner and user must preserve all the labels on the device in a good and legible  

condition throughout the product’s lifetime.

9.3 REPAIRS

Repairs may only be carried out by persons authorised by GCE.

Each product sent for a repair to a person authorised by GCE shall be properly packed.

The reason for a repair must be clearly specified. A short description of the fault and any reference 

to the complaint number will be helpful.

Some repairs concerning the replacement of damaged or missing components may be 

carried out by the product’s user. The following components may only be replaced:

• 

Cannula 

• 

Batteries 

• 

Spiral hose with connection 

• 

Battery compartment cover

Besides batteries, use original GCE parts only.

8

.

9

.

Summary of Contents for ECOlite4000

Page 1: ...EN Page 1 12 ECOlite 4000 GAS SAVER INSTRUCTION FOR USE...

Page 2: ...om the oxygen The dosage and timing of oxygen delivery is controlled by a microprocessor in accordance with the chosen setting Dosage per breath is independent of breathing rate therefore additional o...

Page 3: ...TIONS At atmospheric pressure up to 60 oxygen is harmless 100 oxygen produces a cough after 6 hours and increasing breathlessness after 25 hours PERSONNEL INSTRUCTIONS AND TRAINING According to Medica...

Page 4: ...function from national standards In the event of any problem with connection please contact the distributor the manufacturer Display screen ON OFF button Oxygen inlet Outlet into cannula Control butt...

Page 5: ...clean If the unit is being used with an oxygen supply source with adjustable fl ow rates set the flow rate on the supply source to 4 litres per minute or to the closest setting above Check the cylinde...

Page 6: ...s the button and hold it down Afterwards press the ON OFF button and also hold it down A PRESS KEY sign will appear Press the two buttons marked and simultaneously a LANGUAGE sign will appear Using th...

Page 7: ...ngly Activate the alarm by not breathing in the cannula After approx 1 minute the alarm should set off both by sound and an alert on the display If there is an abnormal sound noise gas leakage or any...

Page 8: ...void wasting oxygen 6 4 2 DISCONNECTING THE DEVICE Before disconnecting the ECOlite 4000 from the gas source Close the oxygen supply from the cylinder reservoir disconnect the connector of the spiral...

Page 9: ...At the end of the product s life time 10 years at most the ECOlite 4000 device must be withdrawn from service To prevent reuse the owner shall make the ECOlite 4000 impossible to use or return it to...

Page 10: ...c Area EEA This mark on the product indicates compliance with Directive 93 42 EEC relating to medical devices Class IIa Functional performance Sensitivity 0 13 cm of H2O Setting Manual automatic Trigg...

Page 11: ...eter Use by Outlet parameter Take back equipment for recycling Do not dispose equipment into unsorted municipal waste WARRANTY The Standard Warranty period is two years from date of receipt by the GCE...

Page 12: ...Asia The company operates 15 subsidiaries around the world and employs more than 850 people GCE Group includes four business areas Cutting Welding Process Applications Medical and High Purity Today s...

Reviews: