5/52
EN
FOR TERMINAL UNIT MEDIUNIT
A
B
C
D
E F G H I
J
K
A - GAS SPECIFIC PLUG
Plugs are part of the button designed in colours specified to the gas for
which TU is designed for.
B - BUTTON
Button serves for disconnection of the medical device connected in the
TU.
C- COVER - RECESSED OR EXPOSED
Recessed cover or exposed cover are basic external part of the complete
TU.
D - QUICK COUPLER (QC)
Quick coupler is used to connect the medical devices into the TU. Quick
coupler allows only the gas specified device for which also the TU is de-
signed for. Quick coupler can be only connected into the housing which is
designed for the same type of gas.
E - SCREWS M5X12
TU includes 3 pcs of screws M5x12 to be used for fixing the housing in the
installation box.
F - O-RING
Housing included an o-ring as a basic sealing element between housing
and quick coupler.
G - CHECK VALVE OF TU
Included in every TU is a check valve that automatically closes the housing
if nothing is connected to the quick coupler.
H - HOUSING WITH AXIAL OR RADIAL INLET
Housing is an in-built part of installation with two general variants:
•
Housing with radial inlet connection
•
Housing with axial inlet connection
I - LOCKING PLATE
Locking plate secures the housing with quick coupler in the installation
box.
J - INSTALLATION BOX
Installation box is used for installation of TU in or on the wall.
K - PLUG
The plug prevents contamination of the installation box.