Then turn the valve case until the locking slot is exactly plumb over the
coupling centre line. Secure the coupling in that position with the locking
nipple (6). Use a 24 mm spanner or special tool 202 200 513 (torque
15 Nm).
•
Test the connection for leaks. Slip the rubber bushing (8) over the
coupling, bevelled end outwards, and press the bushing against the
base of the valve body (2).
•
Insert the name ring (9) into the cover ring (10).
•
Insert the plastic cup (11) into the name ring and slip it over the bushing.
Fasten the plastic cup in the valve body with the two screws (12). Tighten
the screws until the cover ring rests closely against the wall.
•
Fasten the push-release plate (13) to the coupling with the circlip (14).
•
Test the function of the assembled gas outlets by connecting a male
quick coupling for the respective gas.
•
Reject enclosed details intended for exposed assembly.
Keep all equipment, also hands, tools, working clothes clean and free
from grease or oil.
6.1.4. RECESSED INSTALLATIONI MEDIUNIT
J
L
H1
I
E
H2
K
4x Holes (O)
•
Drill a hole of diameter 60 - 65 mm into the wall for installation of the
installation box (J).
•
For axial housing (H2): drill a hole for axial inlet connection into the bot-
tom (marked) of the installation box (J). Seal the radial hole with plug (K)
to protect installation box (J).
For radial housing (H1): continue by the next point.
•
Fix the installation box (J) in the prepared hole with four screws (L).