MCA166-USB
MCA Stabilisation Commands:
Command name
CMD_SET_STABILISATION
Description
The command sets the peak stabilisation.
fl:
0
turns stabilisation off
1
turns stabilisation on
rb+3<fl<re-3
stabilisation to channel fl
rb:
Peak ROI begin (LLD<= begin < end)
re:
Peak ROI end (begin < end <= ULD, (end-begin) < 250)
Command syntax
Format
integer
integer
integer
long
integer
Parameter
preamble
command
fl
rb
re
end flag
Byte String (HEX)
A5
5A
4D
00
fl
l
fl
h
rb
l
rb
h
re
l
re
h
B9
9B
Used setup value
channels, lld, uld
Affected setup value
stab_state, stab_roi_begin, stab_roi_end
MCA state
STATE_READY
Command name
CMD_SET_STAB_PARAM
Description
The command sets the stabilisation parameters.
st:
time interval (default 10)
sa:
area (default 25000)
Command syntax
Format
integer
integer
integer
long
integer
Parameter
Preamble
command
st
sa
end flag
Byte String (HEX)
A5
5A
67
00
st
l
st
h
sa
l
sa
l
sa
l
sa
h
B9
9B
Used setup value
-
Affected setup value
stab_time, stab_area
MCA state
STATE_READY
66