ULTRATHERM 908i
gbo Medizintechnik AG 2004
Version 1.6
The following buttons are assigned to the corresponding applicators:
Button Applicator
Symbol
6
Coil applicator Minode,
∅
8 cm
7
Coil applicator Monode,
∅
14 cm
8
Coil applicator Diplode,18 x 39 cm
9
Capacitor-field applicator; capacitor-field applicator,
∅
4.2 cm elastic, 8x12 cm
10
Capacitor-field applicator; capacitor-field applicator,
∅
8.5 cm elastic, 12x18 cm and larger
11
Capacitor-field applicator;
∅
13 cm
As mentioned above, any programming is only possible with the dosage display
(15)
showing
„0".
WARNING:
The coil applicator
MONODE
must not be used in the pulsed mode of operation!
3.5
Connection of the applicators
Applicators are electrically connected to the sockets
(43).
Insert the plug pins into the socket up to the stop!
Please, observe that in any case only the connectors of the capacitor-field applicator cables or the
connectors of the coil applicator cable are plugged into sockets
(43)
.
3.6
Treatment time
Set the treatment time with buttons
(12)
, increasing, or
(13)
, decreasing. The actual value is
shown on display
(14)
. Pressing the buttons for a short time will set the time in 1 -minute
intervals, pressing them continuously will set the time in 5-minute intervals. When both buttons
are simultaneously pressed the time is reset to „O". The maximum time that can be set is 30
minutes. Changing the time during the treatment is also possible.