Behavior in case of failures
The following failures will be indicated by the
TESI
table both optically and acoustically. Most of them can
be remedied by following the instructions on the display. As a rule, the following applies:
1.
The
TESI
table will not execute any incorrectly entered functions or modifications entered by pressing
on the Touch Screen.
2.
The acoustic error signal will sound.
3.
An error message will appear on the display.
5.1
Notes
Note – Touch the screen with a single finger only. Don’t slide over the screen.
Possible reason: The Touch Screen was touched at several different points or the fingers made a sliding
movement across the screen. The Touch Screen must not be operated in this way.
Suggestion:
1.
Please note that the Touch Screen will only work properly if it is touched only by a single finger at a
single point. If more than one finger or the ball of the thumb are placed on the active surface, this
will trigger the error message shown above.
2.
Confirm the message with the
Button (OK)
.
Note – Time credit of chip card expired. Please recharge your card for future use.
Possible reason: The time credit of the card is expired.
Suggestion:
Recharge time account on the card.
5.2
Warnings
Warning – The stepped bed has been stopped! It should only be moved without load.
Possible reason: The stepped bed can only be moved when not under excessive load.
Suggestion:
1.
The patient must remove his feet from the stepped bed.
2.
Touch the
Button (OK)
.
3.
Move the stepped bed into the desired position.
Warning – The carriage motor is blocked and was stopped.
Possible reason: The carriage cannot be moved and was therefore stopped for safety reasons.
Suggestion:
1.
Remove all unnecessary loads from the cover.
2.
Confirm the message by touching the
Button (OK)
and try again to move the carriage by touching the
Touch Screen.
3.
If the message appears again, contact the Service Center.