Transport and assembly
To place the unit, each plane surface is appropriate. Please take care of having enough space around the
device to reach the power switch comfortably and possibly be able to pull off the power cable. Keep a wall
distance of at least 20 cm. For fixing the device please use the brakes of the rollers.
The device should not be placed in front of radiators or radiant. The device is not made for outdoor
operation.
The
TESI
Relax
corresponds to the regulations of EN 60601-1. It is a device of protection class I. Within
the scope of the Medical Device Directive (MPG) the
TESI
Relax
belongs to class IIb (please observe also
Chapter 7 Warnings and safety precautions).
Please check all components to be in external integrity before use.
Warning!
The unit is not designed to be operated in places with the inherent risk of explosions. If
it is used in dangerous areas of anesthesia departments, the possibility of an explosion
cannot be excluded.
If the patient and/or the patient cable is directly exposed to a radiator of a medical
device for high frequency heat therapy, damage of the device or danger to the patient
cannot be excluded. As a rule, a clearance distance of 2 - 3 m is sufficient.
Do not pierce the heated cover.
In case of any visible operational disturbances, please contact gbo Medizintechnik AG
or one of the service agencies authorized by gbo Medizintechnik
AG immediately .
2.2
Connection and switch-on
The
TESI
Relax
has been set to be connected to supply voltages of 100 - 240 V. Irrespective of the
adjusted mains voltage, the device is appropriate for power frequencies of 50 to 60 Hz.
Connect the device with the mains cable to a socket with protective ground. The protective ground must
work correctly.
The
TESI
Relax
is switched on by the mains switch at the front of the head piece (13). By this
arrangement an erroneous or unintended disconnection of the device during normal operations shall be
avoided.
After switching on the
TESI
Relax
, an automatic check of all functions will be carried out.