gbo Medizintechnik AG
Version 1.4
8. In the detailed view you see some additional information:
TIME: 59:22
U: 12.5
V
R: 75 V/A
I: 165
mA
P: 2062 mW
IDX: 55
9. The treatment stops after the treatment timer reaches zero. You can also stop or
interrupt the treatment by decreasing the intensity to zero.
4.3 Detail view
In the detail view, the following values are shown:
Display data
Meaning
TIME:
remaining time of the actual therapy.
U:
released voltage.
R:
Patient’s resistance.
I:
released current.
P:
released power.
Some information on the detailed view:
In the detailed view you can recall treatment values like shown in the table above.
The time shown is the treatment time and it is decreased in steps of seconds. When
the treatment timer reaches zero, the unit stops the treatment and the intensity is set
to zero.
The values for voltage (U) and current (I) can be used as an indicator for the
following therapies. We are not able to give information about the absolute values
of voltage and current because each patient reacts differently sensible to the
current.
The patient’s resistance (R) gives information about the electric conductivity of the
system. The resistance should fit to the area of 30 – 150 Ohms. The reason for a
resistance greater than 150 Ohms may be:
No Aloe Vera was used to moisten the electrodes
Bad contact because of oil and/or creme on the skin
No contact of the distribution cable to one of the electrodes
The released power (P) is the product of voltage and current and gives some
information about the released energy for one treatment.
*) All values shown on the display are examples!