Temperature setting:
In this interface,
press
◀
. ▲. ▼ key to adjust the value in the (SV) window, set the required
temperature, and after the adjustment, wait for the number to no longer change,
and the temperature can be used for measurement after 30 minutes.
(5)
The blackbody furnace has been tested in operation when it leaves the
factory.
All parameters of the intelligent temperature controller have been optimized.
It is forbidden to modify the parameters by yourself when not necessary.
(6)
After use, cut off the power.
3.
Precautions for using blackbody
Before use, please check if there is enough space around you. The bottom and back of
the device is a heat dissipation area. When using it, pay attention to cleaning up
foreign objects that are easily sucked in to prevent damage to the device!
The black body furnace can be used in a laboratory, a measurement room or a
production site, and it should be placed horizontally on a table during use. When
setting the instrument to be calibrated, it shall be kept at a specified distance from the
bullseye of the black body furnace. At the same time, the instrument calibration
system and the black body radiation surface must be on the same axis.