SUSIE
®
S1001 |
User Guide
Using SUSIE® S1001
| 21
2. Locate the Luer-Lock port
on simulator’s left posterior
shoulder
3. Connect the Luer-Lock fitting
on the end of the modified BP
cuff extra branch to the Luer-Lock port on her left shoulder.
Calibrate the BP cuff before measuring blood pressure.
Blood Pressure Calibration
Before starting the calibration process, place the blood pressure cuff on
the simulator as it would be placed as instructed above.
To calibrate the blood pressure feature using OMNI tablet:
1. Tap the gear icon on the
upper-right corner of the
screen.
2. Select Setup then Calibration
3. Click Blood Pressure Cuff
and click next.
4. Select “Next” in the
calibration window to begin