ORIUS ™ 833 Use
r
’s Guide Series Rev 3
2-3
Figure 2-2
(a) Camera Menu
(b) Camera configuration
•
Select Configure Camera in the Camera menu (Figure (2-2a).
•
In the Camera configuration dialog box (Figure 2-2b), you can change
the name of the camera by typing in the Name field.
•
Image rotation can be specified in the Rotation pull-down menu.
•
The shutter state should only be setup IF the camera was configured
with the optional shutter for this camera
2.3 Dark Correction
•
After the camera has cooled to its operating temperature of 10°C, a
calibration of the bias or thermal counts is required. This is often
referred to as the “dark image”.
Quick tip:
A dark reference image set must be taken before preparing a
gain reference.
•
To prepare a dark reference image obscure the camera, either by
lowering the TEM screen, retracting the camera or turning off the
TEM beam.
•
Under the camera menu select “Prepare dark reference”.
•
Wait about 15 minutes for the procedure to complete.
•
It should not be necessary to repeat this more that once per week and
may be good indefinitely.
Summary of Contents for ORIUS SC 200D
Page 40: ...ORIUS 833 User s Guide Series Rev 3 IV Figure A4 Camera Pneumatic Interface Details...
Page 41: ...ORIUS 833 User s Guide Series Rev 3 V Figure A5 Water lines...
Page 45: ...ORIUS User s Guide Series Rev 3 IX Appendix 3 Camera specifications...
Page 47: ...ORIUS 833 User s Guide Series Rev 3 XI Appendix 4 CE certification...