ORIUS ™ 833 Use
r
’s Guide Series Rev 3
2-9
Figure 2-11
High-resolution image of a crystalline sample
•
Choose “Calibrate image from Diffractogram…” under the
“Microscope” menu (Figure 2-5). A diffractogram will be calculated
after placing a FFT ROI on the image (Figure 2-12).
Figure 2-12
Diffractogram
•
Follow the on screen instructions (Figure 2-13)
•
A red line will appear on the diffractogram.
•
Position the endpoints of the red line on two symmetrical diffraction
peaks (Figure 2-12).