11.5.
How to set alarm in Alarm mode
After password has been confirmed, the detector goes into
level1 mode.
Select ALARM mode by touching “
↑
” or “
↓
” switch.
At this moment, if function key is touched, the detector will go
into submenu of ALARM MODE Level2.
This mode is to configure reset type of alarm. By touching “
↑
”
or “
↓
” switch, the type will be alternatively changed between
“AUTO” and “MANUAL”.
In “AUTO” mode, alarm will be reset automatically. In “MANUAL”
mode, alarm will be reset only when reset switch turns on.
This mode is to configure Relay drive type of alarm. By touching
“
↑
” or “
↓
” switch, the type will be alternatively changed
between “DE-ENERGIZED” and “ENERGIZED”.
“DE-ENERGIZED” mode is inactive ENERGIZER function and
“ENERGIZED” mode is active ENERGIZER function.
This mode is to configure Relay drive type of FAULT. By
touching “
↑
” or “
↓
” switch, the type will be alternatively
changed between “DE-ENERGIZED” and “ENERGIZED”.
“DE-ENERGIZED” mode is inactive ENERGIZER function and
“ENERGIZED” mode is active ENERGIZER function.
This mode is to set the operational direction for Alarm1.
“INCREASE” or “DECREASE” will be displayed alternatively
whenever “
↑
” or “
↓
” switch is touched.
In “INCREASE” mode, alarm1 will be activated if the measured
value is same or higher than preset value for Alarm1.
In “DECREASE” mode, alarm1 will be activated if the measured
value is same or lower than preset value for Alarm1.
ALARM1 TYPE
SEL
ALARM1 TYPE
SEL
FAULT RELAY
TYPE
ALARM RELAY
TYPE
ALARM
OPERATING
ALARM
OPERATING
ALARM
PROGRAM