D-Guard
2
33
12.
Calibration
12.1.
Why Calibrate ?
All electro-chemical gas sensing cells have a finite service life. Through reactions with ambient atmosphere that may
contain contaminants, and in some cases, to target gas, the internal electrodes slowly become less active and efficient.
In operating conditions where the cell is not exposed to its target gas (other than for calibration purposes), and the
surrounding environment does not contain volatile organic compounds (VOC), the cell should operate reliably for
around 18 to 24 months.
Regular exposure to non-clean air, or VOC's will shorten the cells lifetime. The reduced lifetime will depend upon the
frequency and nature of exposure to non-ideal ‘clean-air’ conditions. Once a cell has reached the limit of its lifetime, it
must be replaced (see maintenance section).
For Infra-Red type sensors, the process of performance degradation is much slower and is generally related to changes
in the characteristics of the optical components used to measure the gas concentration. Occasionally, detector
operation in a harsh environment can cause stress to the optical components causing a shift in its performance.
For all sensor types, as a result of its aging mechanism, gas proportional electrical outputs will degrade from the levels
they were when new. This signal degradation can be observed as a baseline measurement shift, either positive or
negative. Similarly, the cell’s sensitivity to gas will reduce with a corresponding reduction in output signal for a given
gas concentration.
If left uncorrected, eventually the detector will no longer operate reliably in detecting its intended gas.
In order to maintain accurate and reliable operation, the detector must be calibrated at regular intervals. Calibration
will return the detector to a correct zero gas baseline and re-adjust operation to compensate for the reduced cell
sensitivity.
Some environmental standards mandate that detectors must be calibrated regularly if the monitoring system is to
remain compliant with its rules. Gastech recommends calibration at 6 monthly intervals for a detector operating under
normal conditions.
12.2.
Preparing for calibration
The method for calibrating the detector has similarities to bump testing. See sections 9.3 to 9.5.
To ready the detector for calibration, conduct the operations covered in sections 9.3 to 9.5.
Unlike bump testing however, a minimum target gas level of 25% of full scale must be used for calibration to ensure an
accurate final calibration and for the detector to achieve its published performance specifications.
D-Guard
2
prevents zero calibrations from being performed if the apparent gas level is greater than 10% of full scale.
Span calibrations are prevented if apparent gas concentration is less than 20% of full scale .
Summary of Contents for D-Guard2
Page 2: ...D Guard2 1 This Page Left Intentionally Blank ...
Page 65: ...D Guard2 62 23 APPENDIX F 23 1 System Menu Map ...
Page 68: ...D Guard2 65 25 APPENDIX H Product Certificates 25 1 IECEx Product Certificate ...
Page 69: ...D Guard2 66 ...
Page 70: ...D Guard2 67 ...
Page 71: ...D Guard2 68 ...
Page 72: ...D Guard2 69 25 2 ATEX Product Certificate ...
Page 73: ...D Guard2 70 ...
Page 74: ...D Guard2 71 ...
Page 75: ...D Guard2 72 ...
Page 76: ...D Guard2 73 26 APPENDIX I Product Dimensions ...
Page 77: ...D Guard2 74 Doc_Rev 03 12 15 ...