8.
This
next
menu
is
used
to
precisely
calibrate
the
full
scale
loop
current
to
20mA
via
a
trimpot
on
the
back
of
the
unit.
9.
Adjust
the
trimpot
shown
below
until
the
reading
on
the
multimeter
shows
exactly
200.000mV.
This
means
that
the
unit
is
correctly
set
to
20.00mA
at
the
full
scale
point.
10.
Once
any
adjustments
have
been
made,
remove
the
jumper
between
pins
1
and
2,
screw
the
lid
back
together,
close
the
calibration
door
and
normal
operation
will
resume.
Full
scale
adjustment