CMM USER'S MANUAL (E2.51)
46
4.6.1.3 Service Alarms:
There are 14 service alarms represented by 14 bits, each bit represents one alarm. 0=No Alarm,
1=Alarm
Bath
Pressur
e
Bath
Flo
w
Dilutio
n Flow
Cabinet
Temperatur
e
Instrumen
t
air
Pressure
Cabine
t
Cooler
Pum
p
Zero
Chec
k
Error
Hg0 Span
Calibratio
n Error
Hg0
Span
Chec
k
Error
HgCl
2
Span
Chec
k
Error
Hg0
Linearit
y Check
Error
HgCl2
Linearit
y Check
Error
Flow
Alar
m
0
0
0
0
0
0
0
0
0
0
0
0
0
0
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Table 16. Service Alarms
4.6.1.4 Result Valid:
There are 5 result valid alarms represented by 5 bits, each bit represents one alarm. 0=No Alarm,
1=Alarm
Standby
System
Alarm
No
Active
Analysed
Result
Result
Range
Exceeded
Maintenance
0
0
0
0
0
1
1
1
1
1
Table 17. Result Valid
4.6.2 Calibration files
The calibration file has its own path and file name for saving in ‘Program settings’. The calibration file
contains the results of zero calibration, span calibration and linearity check (see Figure
28
). The
calibration file contains the following columns:
-
The date, start and end time of the calibration or check performed
-
Serial number of the analyzer
-
Mode of calibration (0=Manual 1=Automatic)
-
The type of adjustment or check performed
-
Status of performed calibration (Pass, Fail)
-
The measured Hg
0
concentration at zero level
-
The measured Hg
0
concentration at span level
-
The measured HgCl
2
concentration at span level (CMM AutoQAL)
-
The measured Hg
0
concentration at linearity check levels
-
The measured HgCl
2
concentration at linearity check levels (CMM AutoQAL)
-
Target concentration
-
Error limit as fraction of full range
-
Spread value of HgCl2 span check calculations
-
Error limit for Spread value of HgCl2 span check as fraction of full test range