CMM USER'S MANUAL (E2.51)
15
2.4 Installation
Improper handling of any components inside the Gasmet CMM may damage the equipment. Thus,
adjusting any components inside the Gasmet CMM voids warranty. Be always very careful if you move
the instrument.
2.4.1 Explosion protection
For your own safety, the Gasmet CMM must not be used in hazardous areas. The
standard cabinet is not explosion proof.
Gasmet Technologies Oy takes no responsibility if the Gasmet CMM is used in
hazardous areas.
2.4.2 Aggressive condensate and temperature
Aggressive condensate may have formed. Wear protected glasses and proper protective clothing.
Wear protective gloves, because surfaces’ temperatures can be high.
The Gasmet CMM must be mounted vertically. An extreme inclination of the system cabinet may
interfere with the operation of the analyzer. The installation location must preferably be free of strong
vibrations, as well. It is recommended that the instrument is operated in the following environmental
conditions:
·
0 - 40 °C, optimum 15 - 25
°
C, non-condensing
·
< 90 % relative humidity at 20
°
C, non-condensing
The ambient temperature of installation location must preferably be stable. The Gasmet CMM cabinet
includes air conditioning unit. Never block the air circulation or use The Gasmet CMM in dusty
environments. Dust can block the air circulation and penetrate into the cabinet damaging analyzer’s
electronics.