FlexCell with inserted gas diffusion electrode between the silicone sealings
FlexCell with inserted metal sheet between the O-rings
When screwing the cell components together, make sure that the screws are tightened
diagonally to each other one after the other. There must be no gap between the
components. Tighten the screws only hand-tight, especially when using the flat gaskets.
Leakage is usually caused by unevenly tightened screws and not by too weakly tightened
ones.
If this is the case, you can now screw the contact screws back in with the socket spanner
and hand-tighten them so that you contact the electrode.