6.3 Operation of the SF
6
-Aciditor
For operating the SF
6
-Aciditor the following instructions and information are needed.
The following steps explain how to operate the SF
6
-Aciditor in detail.
1. Turn on the SF
6
-Aciditor using the
“Power”-switch on the front panel of the
device and wait until the main screen is
displayed (see picture).
2. Connect a sampling bag or a recycling
system at the “Gas Outlet” on the back of
the SF
6
-Aciditor (optional).
3. Connect the SF
6
-Aciditor (Gas Inlet) with the delivered hose (equipped with self-
sealing quick-connectors) and appropriate adapters to the SF
6
-filled compartment.
4. Open the valve at the SF
6
-filled compartment.
Information!
The SF
6
-Aciditor has no internal valve. If there is an open connection to an
SF
6
-filled compartment, the gas passes through the device with 20 L/h. This,
independently whether the device is turned on or off !!