36
6.4.2
Defining the automatic measurement schedule
Carefully plan your Interval schedule, consider the minimum time required
for a measurement/calibration and consider the implemented hierarchy of
priorities in case of coincidence: Cleaning has priority over Calibration and
Calibration has priority over Measurement.
During a measurement device component values are dynamically modified.
These values are displayed in the scope view.
Any cleaning process must be completed ~2 hours prior to any
measurement so that the internal components can cool down to the
required temperatures.
This must be taken into account also if cleaning cycles are introduced in
the interval schedule
Before every measurement and before every calibration the RIP and the
default parameters are tested. In case this test fails the measurement is
invalidated or the calibration is skipped.
1
Select mode button:
Start
›
INTERVAL
.