35
6.4
Workflow Running automated Measurements
6.4.1
Introduction to the Interval Mode
In the interval mode scheduled measurements and calibrations are performed in a
continuous and predefined sequence. The latest release of the GC-IMS-Odor
Firmware provides a full flexibility to define the measurement intervals, as well
as calibration and cleaning intervals.
All substance calibrations associated with the selected application
mode are modified within the interval mode when calibrations are
activated. Make sure that an appropriate calibration gas source is
connected to the Testgas In socket at the rear of the housing.
A calibration only works within certain margins around the factory
calibration. When it varies too much from the factory calibration the
user is informed and the results of the calibration process are
discarded.
6
This Calibration measure-
ment failed because the
Correction Facture was out of
the
defined
acceptable
range.